Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corporation: Terumo Backplate Replacement CXFVR 1500 Holder The Flexible venous Reservoir is mounted on the Backplate/Holder and secures the flexible reservoir by strategically placed mounting arms. Additionally, there are three port supports the user incorporates to position the lines that enter and exit the venous bag. The Backplate/Holder is designed to allow the user to move the port support location to accommodate different tubing circuit designs to permit mounting flexibility.

Recall number
Z-0071-2013
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Backplate Replacement CXFVR 1500 Holder The Flexible venous Reservoir is mounted on the Backplate/Holder and secures the flexible reservoir by strategically placed mounting arms. Additionally, there are three port supports the user incorporates to position the lines that enter and exit the venous bag. The Backplate/Holder is designed to allow the user to move the port support location to accommodate different tubing circuit designs to permit mounting flexibility.
Status
Terminated
Initiated
2009-10-02
Posted
2012-10-16
Terminated
2013-01-29
Reason
Backplate reservoir holders were manufactured with an incorrect configuration of connector supports. The devices were manufactured and configured with two 3/8 inch and one inch connector supports instead of one 3/8 inch and two inch connector supports (See location of supports as # 7 and # 8 on backplate attachment). The backplates were manufactured as described in manufacturing instructions a
Root cause
Labeling design
Action
The firm, Terumo, interviewed the Sr. Product Manager from the Marketing Department of the company on August 29, 2012. The interview included the description of the product, problem and action taken. Replacement connectors were provided to users and all units were corrected. If you have any questions, call 734-663-4145 or Manager, Regulatory Affairs at 734-741-5576.
Quantity
33 units
Distribution
Worldwide distribution: USA (nationwide) including states of: CA, MI, NC, NJ, PA, and WI;and country of: EUROPE.
Lot, serial or model codes
No lot/serial number; Catalog number: PN 816280
510(k) numbers
not on file
PMA numbers
not on file
Product code
KRI
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register