Skip to the content

FDA Medical Device Recalls (openFDA)

Leica Microsystems, Inc.: Leica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated Product Usage: The intended use of this device is a stainer for laboratory applications and intended for the preparation of histological and cytological specimens.

Recall number
Z-0064-2019
Recalling firm
Leica Microsystems, Inc.
Product
Leica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated Product Usage: The intended use of this device is a stainer for laboratory applications and intended for the preparation of histological and cytological specimens.
Status
Terminated
Initiated
2018-06-13
Posted
not on file
Terminated
2019-03-18
Reason
These devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user.
Root cause
Device Design
Action
On approximately 06/13/2018 sent a Field Action Notices to consignees. Instructions included to make arrangements with their Leica Service Engineer to correct affected devices, ensure all appropriate personnel are made aware of the recall, and complete and return the Field Action Notice Return Response Form to confirm receipt and understanding of the recall notice.
Quantity
67 units
Distribution
Distributed nationwide to AZ, CA, CO, KS, MA, MH, MO, NJ, OH, WI. Distributed worldwide to CANADA, CHINA, GERMANY, GREAT BRITAIN, JAPAN, SWITZERLAND
Lot, serial or model codes
Serial Numbers: 70-78, 111-112, 120-175
510(k) numbers
not on file
PMA numbers
not on file
Product code
KPA
Firm FEI
3000210133
Firm city
Buffalo Grove
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register