Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Baxter''s Patient Care System (BPCS) software, Pharmacy Management - Inpatient - product code 2M5013 and Point of Care - product code 2M5014; Baxter Healthcare Corporation, 1 Baxter Parkway, Deerfield, IL 60015, USA, made in Canada

Recall number
Z-0061-06
Recalling firm
Baxter Healthcare Corp.
Product
Baxter''s Patient Care System (BPCS) software, Pharmacy Management - Inpatient - product code 2M5013 and Point of Care - product code 2M5014; Baxter Healthcare Corporation, 1 Baxter Parkway, Deerfield, IL 60015, USA, made in Canada
Status
Terminated
Initiated
2005-08-29
Posted
2005-10-15
Terminated
2008-02-06
Reason
If an administrative user modifies any of the seven pre-defined Monitoring Parameters when they are attached to a medication in the drug file, a Point of Care end user will be unable to enter monitoring parameter results on the handheld.
Root cause
Other
Action
Urgent Device Correction recall letters dated 8/29/05 were sent to the direct accounts on the same date via first class mail. The letter describes the issue with the modification or creation of a Monitoring Parameter, and informed the accounts that modifying the Monitoring Parameters could result in a potential for a patient receiving an incorrect medication or an incorrect medication dosage. The accounts were advised that administrative end users should never modify the seven pre-defined Monitoring Parameters or create new Monitoring Parameters.
Quantity
5 units
Distribution
Missouri, Massachusetts, Vermont, California and Canada.
Lot, serial or model codes
all units of software versions 4.0 (66) and 4.1 (77)
510(k) numbers
not on file
PMA numbers
not on file
Product code
LNX
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register