FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: EVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera heads, endoscopes, light sources, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation
- Recall number
- Z-0056-2025
- Recalling firm
- Olympus Corporation of the Americas
- Product
- EVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera heads, endoscopes, light sources, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation
- Status
- Open, Classified
- Initiated
- 2023-09-14
- Posted
- 2024-10-11
- Terminated
- not on file
- Reason
- Certain serial numbers of the CV-190 do not start up properly because parts that deviated from the specification were assembled into the power supply unit.
- Root cause
- Nonconforming Material/Component
- Action
- An URGENT: MEDICAL DEVICE FIELD CORRECTIVE ACTION notification letter dated 9/14/23 was sent to customers. Action steps to be taken by the end user: Olympus has determined based upon our distribution records that your facility is in possession of one or more affected CV-190 with a serial number shown in Attachment 1. Olympus requires that you take the following actions: 1. Inspect your inventory for the referenced products and identify any devices with the model number and serial number specified in Attachment 1. Cease use of products and quarantine any affected products. Please check all areas of the hospital to determine if any of these products remain in inventory. 2. Please contact our Olympus Repair Center at 1-800-848-9024, option 3, to arrange for a service to exchange the power supply unit in affected devices. Olympus regrets any inconvenience caused and fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at 647-999-3203 or Cynthia.Ow@Olympus.com.
- Quantity
- 45 units
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- UDI-DI: 04953170298622; Serial Numbers: 7384484 7384422 7384487 7384478 7384479 7384442 7384441 7384485 7384451 7384455 7384481 7384470 7384424 7384423 7384489 7384531 7384516 7384492 7384512 7384543 7384483 7384624 7384623 7384497 7384495 7384494 7384533 7384622 7384660 7384654 7384656 7384647 7384658 7384675 7384655 7384704 7384709 7384687 7384706 7384661 7384662 7384703 7384670 7384679 7384678
- 510(k) numbers
- K131780
- PMA numbers
- not on file
- Product code
- FDF
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28