Skip to the content

FDA Medical Device Recalls (openFDA)

Brainlab AG: Ultrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x and 3.1.x - Product Usage: intended to be an intra-operative image guided localization system to enable minimally invasive surgery.

Recall number
Z-0056-2021
Recalling firm
Brainlab AG
Product
Ultrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x and 3.1.x - Product Usage: intended to be an intra-operative image guided localization system to enable minimally invasive surgery.
Status
Terminated
Initiated
2020-08-31
Posted
not on file
Terminated
2021-02-17
Reason
Brainlab Ultrasound Navigation Software does not support the modification of the probe's image width.
Root cause
Software design
Action
The intraoperative Brainlab Ultrasound Navigation Software does not support the modification of the probes image width, an optional feature of the BK Medical Ultrasound System, resulting in an incorrect display of the depth of the ultrasound overlay by the navigation for non-linear ultrasound probes, with the deviation depending on the probe and the difference between the calibrated and used image width.
Quantity
271 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of Illinois, Massachusetts, Texas, California, Idaho, Missouri, Oregon, Connecticut and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Curacao, Czech Republic, Egypt, Finland, France, Germany, Switzerland, Netherlands, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kuwait, Malaysia, Myanmar, Norway, Pakistan, Romania, Slovakia, South Africa, Spain, Sweden, Thailand, Turkey and United Kingdom.
Lot, serial or model codes
Model/catalogue numbers: 22582 IGSONIC BK MEDICAL SOFTWARE 22582A ULTRASOUND INTEGRATION SW (BK) 22582B ULTRASOUND INTEGRATION SW 22583 LICENSE UPDATE TO IGSONIC SW FOR BK MED 26506 ULTRASOUND SW BK800 SUBSCRIPTION 26506A ULTRASOUND SW BK SUBSCRIPTION 26508A ULTRASOUND SW BK UPDATE  UDI: The GTIN for Ultrasound Navigation 1.0 is: 04056481140137 The GTIN for Cranial/ENT Navigation 3.1 is: 04056481132439
510(k) numbers
["K092467"]
PMA numbers
not on file
Product code
HAW
Firm FEI
3002619595
Firm city
Munich
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register