Skip to the content

FDA Medical Device Recalls (openFDA)

Kinova Inc: KINOVA Jaco assistive robot. Model Numbers: PJ 0000 0001, PJ 0000 0012, PJ 0090 0001, PJ 0090 0006, KR MJ2 0001. Jaco replaces the function of the upper limb on one side of the body.

Recall number
Z-0049-2025
Recalling firm
Kinova Inc
Product
KINOVA Jaco assistive robot. Model Numbers: PJ 0000 0001, PJ 0000 0012, PJ 0090 0001, PJ 0090 0006, KR MJ2 0001. Jaco replaces the function of the upper limb on one side of the body.
Status
Open, Classified
Initiated
2024-08-22
Posted
2024-10-11
Terminated
not on file
Reason
Potential installation issue that could result in a fire hazard in the event of an interaction between a damaged Jaco arm and a wheelchair exhibiting electrical leakage.
Root cause
Device Design
Action
On 07/22/2024, Kinova notified consignees via UGENT: MEDICAL DEVICE CORRECTION letter. Consignees were instructed contact Kinova immediately for an assessment, ensure that if the wheelchair is in contact with a damaged part of the Jaco arm in any position at any time, to unplug the Jaco arm from the power source immediately and contact Kinova, and read the updated user guide on Kinova's website. Consignees were also requested to complete and return the acknowledgement and response form. Distributors were instructed to communicate the recall to all Jaco users, confirm receipt and understanding of the recall communication, maintain a detailed list of users, provide the user's contact details to Kinova, and perform an assessment for each user that response to the notification. Distributors were also instructed to complete and return the acknowledgement and response form.
Quantity
949 units
Distribution
Worldwide - US Nationwide distribution in the states of California, Missouri, North Carolina, Ohio and the countries of Australia, Canada, China, Czechia, France, Germany, Netherlands, Norway, Portugal, Switzerland, Great Britain, Northern Ireland.
Lot, serial or model codes
Model Numbers: PJ 0000 0001, PJ 0000 0012, PJ 0090 0001, PJ 0090 0006, KR MJ2 0001. All lot numbers.
510(k) numbers
not on file
PMA numbers
not on file
Product code
IQZ
Firm FEI
3009538853
Firm city
Montreal
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register