Skip to the content

FDA Medical Device Recalls (openFDA)

Edap Technomed Inc.: Ablatherm(R) Integrated Imaging, High Intensity Ultrasound System For Prostate Tissue Ablation Transrectal high intensity focused ultrasound (HIFU) ablation of prostate tissue

Recall number
Z-0049-2018
Recalling firm
Edap Technomed Inc.
Product
Ablatherm(R) Integrated Imaging, High Intensity Ultrasound System For Prostate Tissue Ablation Transrectal high intensity focused ultrasound (HIFU) ablation of prostate tissue
Status
Terminated
Initiated
2017-08-14
Posted
not on file
Terminated
2020-08-04
Reason
The US FDA has requested the optional energy treatment settings, "medium" and "low" (Software Protocols 02-Medium and 03-Low) be removed from all Ablatherm Integrated Imagining devices in the U.S. until supporting clinical data can be submitted and evaluated by FDA.
Root cause
Other
Action
The firm, EDAP TMS, company contacted all affected customers by phone during the week of August 14, 2017 and followed up with an " URGENT FIELD SAFETY NOTICE" letter dated 8/22/2017 delivered by email on 08/29/2016. The letter described the product, problem and actions to be taken. The customers were instructed as follows: do not use the "medium" or "low" energy treatment power settings in the meantime.The devices will be serviced at the consignee to remove Protocols "02-medium" and "03-low". If you have any questions regarding this notification, or experience any issues with your device, please contact EDAP TMS Customer Care Center at (512) 852-9685 or via email at service@edap-usa.com at any time, 24 hours a day, 7 days a week.
Quantity
12 units
Distribution
US Distribution to states of: NY, FL CA, NC, TX and NJ.
Lot, serial or model codes
Serial Numbers: AB127r, AB206r, AB220r, AB254r, AB200r, AB214r, AB223r, AB266, AB204r, AB217r, AB224r, AB267
510(k) numbers
["K153023"]
PMA numbers
not on file
Product code
PLP
Firm FEI
1287394
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register