FDA Medical Device Recalls (openFDA)
Edan Instruments, Inc.: ECG Cable Part Number: 01.57.471017 for use with the SE-1 series, SE-3 series and SE-601 series Smart Electrocardiograph ECG cable is an accessory of the following machines: SE-1 series, SE-3 series and SE-601 series electrocardiograph (ECG). The EDAN ECG is a diagnostic and prescription device only intended to be used in hospitals or healthcare facilities by qualified doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help medical professionals to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. ECG cable is packed in plastic bag, as an accessory putting together with ECG machine in the package box.
- Recall number
- Z-0041-2012
- Recalling firm
- Edan Instruments, Inc.
- Product
- ECG Cable Part Number: 01.57.471017 for use with the SE-1 series, SE-3 series and SE-601 series Smart Electrocardiograph ECG cable is an accessory of the following machines: SE-1 series, SE-3 series and SE-601 series electrocardiograph (ECG). The EDAN ECG is a diagnostic and prescription device only intended to be used in hospitals or healthcare facilities by qualified doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help medical professionals to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. ECG cable is packed in plastic bag, as an accessory putting together with ECG machine in the package box.
- Status
- Terminated
- Initiated
- 2010-09-10
- Posted
- 2011-10-12
- Terminated
- 2011-10-13
- Reason
- The 3 leads of the cable, RL, LA and LL were labeled in wrong sequence. The left arm (LA) is mislabeled as right leg (RL). The left leg (LL) is mislabeled as left arm (LA). The right leg (RL) is mislabeled as left leg (LL). There have been no reports of injury as a result of this issue.
- Root cause
- Vendor change control
- Action
- Edan Instruments, Inc. sent an Urgent - Medical Device Correction letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that wrongly labeled cable may lead to incorrect ECG waveforms during ECG test, customers were reminded to review previous diagnoses made from this cable to mitigate any inappropriate recordings or treatment made. Customers were instructed to replace the affected ECG Cables once they receive the new ECG Cable, and collect all the affected ECG Cables and then destroy and dispose of them according to local regulations. Questions or concerns were directed to the following numbers: Oceania and Asia: +86 (755) 2689 8322 North America, South America: +86 (755) 2689 9914 Europe: +86 (755) 2669 4684
- Quantity
- 615 Ccables
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Dominica, Ecuador, Colombia, Honduras, Peru, Mexico, El Salvador, Guatemala, Venezuela, Uruguay, Chile, Panama, Nicaragua, Ecuador, Australia, Philippines, Malaysia, Indonesia, Chinese Taipei, & Spain
- Lot, serial or model codes
- Lot Numbers: P09100310 P09100406 P09100428
- 510(k) numbers
- ["K091513"]
- PMA numbers
- not on file
- Product code
- DPS
- Firm FEI
- 3003971136
- Firm city
- Shenzhen
- Firm state
- Guangdong
- Firm country
- China
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as