Skip to the content

FDA Medical Device Recalls (openFDA)

Optovue Inc.: Reference materials "RETINAL OCT Analysis and Interpretation Method" shipped with iVue System (K091404) and the "Retina Poster" shipped with iVue System (K091404) and RTVue Systems (K101505); Models: iVue-100, RTVue-100; The physical Product is manufactured and distributed by Optovue, Inc. Fremont, CA The product is an optical coherence tomography system indicated for the in vivo imaging and measurement of the retina, retinal nerve fiber layer, optic disk, and for cornea and anterior eye scans as an aid in the diagnosis and management of retinal disease.

Recall number
Z-0024-2013
Recalling firm
Optovue Inc.
Product
Reference materials "RETINAL OCT Analysis and Interpretation Method" shipped with iVue System (K091404) and the "Retina Poster" shipped with iVue System (K091404) and RTVue Systems (K101505); Models: iVue-100, RTVue-100; The physical Product is manufactured and distributed by Optovue, Inc. Fremont, CA The product is an optical coherence tomography system indicated for the in vivo imaging and measurement of the retina, retinal nerve fiber layer, optic disk, and for cornea and anterior eye scans as an aid in the diagnosis and management of retinal disease.
Status
Terminated
Initiated
2012-09-07
Posted
2012-10-09
Terminated
2013-03-04
Reason
Devices labeled for an intended use not included in the existing 510(k) or PMA. Specifically, "The Retina" poster provided with the systems, and the book "RETINAL OCT Analysis and Interpretation Method" by Bruno Lumbroso and Marco Rispoli which are not labeled within 51O(k) cleared product description.
Root cause
No Marketing Application
Action
Optovue Inc., sent a Voluntary Field Correction and Removal letter dated September 6, 2012, to all affected customers. The letter identified the product the problem and the action to be taken by the customer . Please sign and return the attached form indicating your acknowledgement that that the Affected Materials have been returned or destroyed. If you have any further questions please contact Optovue, at 1-866-344-8948.
Quantity
795 units
Distribution
Worldwide Distribution -- USA (nationwide) and Internationally.
Lot, serial or model codes
All units of these models.
510(k) numbers
K091404, K101505
PMA numbers
not on file
Product code
OBO
Firm FEI
3005950902
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register