Skip to the content

FDA Medical Device Recalls (openFDA)

Toshiba American Medical Systems Inc: Toshiba Medical Kalare Fluoroscopic X-Ray System (DREX-KL80) Kalare is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations. It is intended for use by qualified/trained doctors or technologists on both adult and pediatric subjects taking radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing or lying in prone or supine positions.

Recall number
Z-0012-2018
Recalling firm
Toshiba American Medical Systems Inc
Product
Toshiba Medical Kalare Fluoroscopic X-Ray System (DREX-KL80) Kalare is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations. It is intended for use by qualified/trained doctors or technologists on both adult and pediatric subjects taking radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing or lying in prone or supine positions.
Status
Terminated
Initiated
2017-09-12
Posted
not on file
Terminated
2020-11-18
Reason
During an examination images were displayed on the live monitor,but the images were not displayed on the system monitor nor were they saved to the hard disk.
Root cause
Software design
Action
Toshiba planned action to bring defect into compliance: 1. You will issue a Customer Notification Letter to make customers aware of the hazards, what measures will be taken to remedy the defect at no cost to the user, and how to safely use the equipment before corrections can be made. 2. You will install modified firmware on the systems to correct the defect and prevent recurrence. 3. You will provide progress reports on the effectiveness of the plan. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator are to be included in the notification. Thank you for your urgent attention to this matter. If you have any questions regarding this letter please feel free to call (800) 421-1968, or contact your local Toshiba Representative at (800) 521-1968. For further questions please call (714) 730-5000.
Quantity
US - 96
Distribution
USA (nationwide) Distribution
Lot, serial or model codes
Model: HDR-08A
510(k) numbers
K133553
PMA numbers
not on file
Product code
JAA
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register