Welch Allyn Inc Mortara: ELI, Burdick and McKesson brand 280 Resting Electrocardiographs (ELI 280, MLBUR280 and BUR280), Model Numbers: ELI280-DDB-ADAAX, ELI280-BDB-ACAAX, ELI280-CAA-AAFAD, ELI280-BDB-BDFAX, ELI280-BCB-AAAAX, ELI280-DDB-ACAAX, ELI280-DBA-BAFAX, ELI280-DCB-AAAAX, ELI280-DBA-AAFBD, ELI280-AAA-AAEBX, ELI280-BBA-ADFBD, ELI280-DDB-ACFAX, MLBUR280-81X, ELI280-CAA-ACEBX, ELI280-AAA-AAHBX, ELI280-DDB-BDFAX, ELI280-CAA-AAFBT, ELI280-DCD-ADFAD, ELI280-DDB-AACBX, ELI280-JXX-BDFAX, ELI280-CAA-ADCBX, ELI280-BCB-AACBD, ELI280-LDX-ADFBX, ELI280-LDX-ADCBX, ELI280-BDD-ADFAD, ELI280-BDB-AAFBD, ELI280-BDB-ACCAD, ELI280-BDB-ADFAD, ELI280-LDX-ADFBD, ELI280-DDB-AAFBD, ELI280-DCB-ACFAD, ELI280-AAA-BAFAF, MLBUR280-W1X, ELI280-DCB-AAABX, ELI280-AAA-ACAAX, ELI280-ADA-ABFBX, ELI280-BCB-AAFBD, ELI280-BCB-AAFBX, ELI280-CEA-ADFBX, ELI280-CDA-ADABX, ELI280-BDB-AAABX, ELI280-DDB-ACFBD, ELI280-DDB-BCFAX, ELI280-ADA-ADCAX, ELI280-BDB-AACBX, ELI280-BCB-ACAAX, BUR280-81X, ELI280-DCB-AAFBG, ELI280-BDB-AAFBT, ELI280-DCB-AAFAD, ELI280-CAA-ADFBX, ELI280-BDB-AAFBG, ELI280-BDB-AAFBX, ELI280-DCA-ACAAX, ELI280-CAB-ACFBX, ELI280-ADA-ACFAX, ELI280-CAA-AAFAT, ELI280-CAA-ACFBD, ELI280-CDA-ADCBX, ELI280-AAA-ADCBX, ELI280-CAA-AAFBD, ELI280-DCB-BAFBT, ELI280-BBA-AAAAX, ELI280-BBA-AAFAD, ELI280-DBA-AAFAD, ELI280-DCB-AAFBT, ELI280-DBA-ADFAX, ELI280-BBA-AAFBD, MLBUR280-C1X, ELI280-BBA-ADFAX, ELI280-DDD-ADFAD, ELI280-DBA-AAFAX, MLBUR280-W1D, ELI280-BCB-AACBX, ELI280-DCB-AACBD, ELI280-LDX-ADFBG, ELI280-DCB-AACBX, ELI280-DBA-ADFBD, ELI280-DDB-AAFBG, ELI280-DBA-AAABD, ELI280-DCB-AAFBX, ELI280-AAA-AAFBT, MLBUR280-C1D, ELI280-AAB-ADAAX, ELI280-DDB-AAFBT, ELI280-CAA-AACBX, ELI280-CEB-ACFBX, ELI280-CAA-ABFAX, ELI280-DDB-AAFBX, ELI280-DBA-AAAAX, ELI280-BCB-AAFBG, ELI280-BFA-ADCBX, ELI280-LDX-ADABX, MLBUR280-81D, ELI280-BBA-ADFAD, ELI280-CAA-ADHAX, ELI280-CEB-ACFBD, ELI280-AAB-ADCAD, ELI280-BDB-AACBD, ELI280-LDX-ADCBD, ELI280-AAB-ACCBX, ELI280-AAA-ABFBX, ELI280-DAB-ADCAD, ELI280-AFB-ABCBX, ELI280-DCB-AACAX, ELI280-DEB-ACFBD, ELI280-ADA-ACAAX, ELI280-AAB-ADFAD, ELI280-DCB-AAFBD, ELI280-CAA-AAAAX, ELI280-BDB-ACCAX, ELI280-CAA-AAFBX, ELI280-DDB-AAABX, ELI280-DCB-ACAAX, ELI280-CAA-ADFBD, ELI280-CBB-ACCBX, ELI280-DDB-AACBD, ELI280-DDB-ACCAX, ELI280-A, ELI280-DFA-ADCBX, ELI280-BCA-AAAAX, ELI280-DDB-AAAAX, ELI280-E, ELI280-C, ELI280-AAA-AAFBD, ELI280-DFC-ADFAD, BUR280-C1X, ELI280-D, ELI280-BAA-ACCBD, ELI280-AAA-AAAAX, BUR280-W1X, ELI280-B, ELI280-BCB-AAABX, ELI280-BCB-AACAX, BUR280-W1D, ELI280-F, ELI280-BCB-AAFAD, ELI280-BCB-BAFAX, BUR280-81D

Recall number
Z-0011-2022
Recalling firm
Welch Allyn Inc Mortara
Product
ELI, Burdick and McKesson brand 280 Resting Electrocardiographs (ELI 280, MLBUR280 and BUR280), Model Numbers: ELI280-DDB-ADAAX, ELI280-BDB-ACAAX, ELI280-CAA-AAFAD, ELI280-BDB-BDFAX, ELI280-BCB-AAAAX, ELI280-DDB-ACAAX, ELI280-DBA-BAFAX, ELI280-DCB-AAAAX, ELI280-DBA-AAFBD, ELI280-AAA-AAEBX, ELI280-BBA-ADFBD, ELI280-DDB-ACFAX, MLBUR280-81X, ELI280-CAA-ACEBX, ELI280-AAA-AAHBX, ELI280-DDB-BDFAX, ELI280-CAA-AAFBT, ELI280-DCD-ADFAD, ELI280-DDB-AACBX, ELI280-JXX-BDFAX, ELI280-CAA-ADCBX, ELI280-BCB-AACBD, ELI280-LDX-ADFBX, ELI280-LDX-ADCBX, ELI280-BDD-ADFAD, ELI280-BDB-AAFBD, ELI280-BDB-ACCAD, ELI280-BDB-ADFAD, ELI280-LDX-ADFBD, ELI280-DDB-AAFBD, ELI280-DCB-ACFAD, ELI280-AAA-BAFAF, MLBUR280-W1X, ELI280-DCB-AAABX, ELI280-AAA-ACAAX, ELI280-ADA-ABFBX, ELI280-BCB-AAFBD, ELI280-BCB-AAFBX, ELI280-CEA-ADFBX, ELI280-CDA-ADABX, ELI280-BDB-AAABX, ELI280-DDB-ACFBD, ELI280-DDB-BCFAX, ELI280-ADA-ADCAX, ELI280-BDB-AACBX, ELI280-BCB-ACAAX, BUR280-81X, ELI280-DCB-AAFBG, ELI280-BDB-AAFBT, ELI280-DCB-AAFAD, ELI280-CAA-ADFBX, ELI280-BDB-AAFBG, ELI280-BDB-AAFBX, ELI280-DCA-ACAAX, ELI280-CAB-ACFBX, ELI280-ADA-ACFAX, ELI280-CAA-AAFAT, ELI280-CAA-ACFBD, ELI280-CDA-ADCBX, ELI280-AAA-ADCBX, ELI280-CAA-AAFBD, ELI280-DCB-BAFBT, ELI280-BBA-AAAAX, ELI280-BBA-AAFAD, ELI280-DBA-AAFAD, ELI280-DCB-AAFBT, ELI280-DBA-ADFAX, ELI280-BBA-AAFBD, MLBUR280-C1X, ELI280-BBA-ADFAX, ELI280-DDD-ADFAD, ELI280-DBA-AAFAX, MLBUR280-W1D, ELI280-BCB-AACBX, ELI280-DCB-AACBD, ELI280-LDX-ADFBG, ELI280-DCB-AACBX, ELI280-DBA-ADFBD, ELI280-DDB-AAFBG, ELI280-DBA-AAABD, ELI280-DCB-AAFBX, ELI280-AAA-AAFBT, MLBUR280-C1D, ELI280-AAB-ADAAX, ELI280-DDB-AAFBT, ELI280-CAA-AACBX, ELI280-CEB-ACFBX, ELI280-CAA-ABFAX, ELI280-DDB-AAFBX, ELI280-DBA-AAAAX, ELI280-BCB-AAFBG, ELI280-BFA-ADCBX, ELI280-LDX-ADABX, MLBUR280-81D, ELI280-BBA-ADFAD, ELI280-CAA-ADHAX, ELI280-CEB-ACFBD, ELI280-AAB-ADCAD, ELI280-BDB-AACBD, ELI280-LDX-ADCBD, ELI280-AAB-ACCBX, ELI280-AAA-ABFBX, ELI280-DAB-ADCAD, ELI280-AFB-ABCBX, ELI280-DCB-AACAX, ELI280-DEB-ACFBD, ELI280-ADA-ACAAX, ELI280-AAB-ADFAD, ELI280-DCB-AAFBD, ELI280-CAA-AAAAX, ELI280-BDB-ACCAX, ELI280-CAA-AAFBX, ELI280-DDB-AAABX, ELI280-DCB-ACAAX, ELI280-CAA-ADFBD, ELI280-CBB-ACCBX, ELI280-DDB-AACBD, ELI280-DDB-ACCAX, ELI280-A, ELI280-DFA-ADCBX, ELI280-BCA-AAAAX, ELI280-DDB-AAAAX, ELI280-E, ELI280-C, ELI280-AAA-AAFBD, ELI280-DFC-ADFAD, BUR280-C1X, ELI280-D, ELI280-BAA-ACCBD, ELI280-AAA-AAAAX, BUR280-W1X, ELI280-B, ELI280-BCB-AAABX, ELI280-BCB-AACAX, BUR280-W1D, ELI280-F, ELI280-BCB-AAFAD, ELI280-BCB-BAFAX, BUR280-81D
Status
Open, Classified
Initiated
2021-09-03
Posted
not on file
Terminated
not on file
Reason
The devices malfunction under specific operator workflows.
Root cause
Software design
Action
The firm notified its consignees on 9/3/2021 by letter via U PS or via USPS first class for those with P.O. Box mailing addresses. The letter notified customers of the issue, provided clinical mitigations and advised them not to utilize specific workflows until their software is updated.
Quantity
14,630 units
Distribution
Distribution was nationwide, including Puerto Rico. Foreign distribution was made to Afghanistan, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Czechia, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Guatemala, Hongkong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Monaco, Morocco, Netherlands, New Zealand, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, The Netherlands, Trinidad and Tobago, Turkey, United Arab Emirates, and United Kingdom.
Lot, serial or model codes
Software versions 2.1.0-2.3.0, Serial numbers 116280503226 thru 121250000503 manufactured from 01-July-2016 thru 01-Jul-2021.
510(k) numbers
K122073
PMA numbers
not on file
Product code
DPS
Firm FEI
2183461
Firm city
Milwaukee
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register