FDA Medical Device Recalls (openFDA)
Welch Allyn Inc Mortara: ELI, Burdick and McKesson brand 280 Resting Electrocardiographs (ELI 280, MLBUR280 and BUR280), Model Numbers: ELI280-DDB-ADAAX, ELI280-BDB-ACAAX, ELI280-CAA-AAFAD, ELI280-BDB-BDFAX, ELI280-BCB-AAAAX, ELI280-DDB-ACAAX, ELI280-DBA-BAFAX, ELI280-DCB-AAAAX, ELI280-DBA-AAFBD, ELI280-AAA-AAEBX, ELI280-BBA-ADFBD, ELI280-DDB-ACFAX, MLBUR280-81X, ELI280-CAA-ACEBX, ELI280-AAA-AAHBX, ELI280-DDB-BDFAX, ELI280-CAA-AAFBT, ELI280-DCD-ADFAD, ELI280-DDB-AACBX, ELI280-JXX-BDFAX, ELI280-CAA-ADCBX, ELI280-BCB-AACBD, ELI280-LDX-ADFBX, ELI280-LDX-ADCBX, ELI280-BDD-ADFAD, ELI280-BDB-AAFBD, ELI280-BDB-ACCAD, ELI280-BDB-ADFAD, ELI280-LDX-ADFBD, ELI280-DDB-AAFBD, ELI280-DCB-ACFAD, ELI280-AAA-BAFAF, MLBUR280-W1X, ELI280-DCB-AAABX, ELI280-AAA-ACAAX, ELI280-ADA-ABFBX, ELI280-BCB-AAFBD, ELI280-BCB-AAFBX, ELI280-CEA-ADFBX, ELI280-CDA-ADABX, ELI280-BDB-AAABX, ELI280-DDB-ACFBD, ELI280-DDB-BCFAX, ELI280-ADA-ADCAX, ELI280-BDB-AACBX, ELI280-BCB-ACAAX, BUR280-81X, ELI280-DCB-AAFBG, ELI280-BDB-AAFBT, ELI280-DCB-AAFAD, ELI280-CAA-ADFBX, ELI280-BDB-AAFBG, ELI280-BDB-AAFBX, ELI280-DCA-ACAAX, ELI280-CAB-ACFBX, ELI280-ADA-ACFAX, ELI280-CAA-AAFAT, ELI280-CAA-ACFBD, ELI280-CDA-ADCBX, ELI280-AAA-ADCBX, ELI280-CAA-AAFBD, ELI280-DCB-BAFBT, ELI280-BBA-AAAAX, ELI280-BBA-AAFAD, ELI280-DBA-AAFAD, ELI280-DCB-AAFBT, ELI280-DBA-ADFAX, ELI280-BBA-AAFBD, MLBUR280-C1X, ELI280-BBA-ADFAX, ELI280-DDD-ADFAD, ELI280-DBA-AAFAX, MLBUR280-W1D, ELI280-BCB-AACBX, ELI280-DCB-AACBD, ELI280-LDX-ADFBG, ELI280-DCB-AACBX, ELI280-DBA-ADFBD, ELI280-DDB-AAFBG, ELI280-DBA-AAABD, ELI280-DCB-AAFBX, ELI280-AAA-AAFBT, MLBUR280-C1D, ELI280-AAB-ADAAX, ELI280-DDB-AAFBT, ELI280-CAA-AACBX, ELI280-CEB-ACFBX, ELI280-CAA-ABFAX, ELI280-DDB-AAFBX, ELI280-DBA-AAAAX, ELI280-BCB-AAFBG, ELI280-BFA-ADCBX, ELI280-LDX-ADABX, MLBUR280-81D, ELI280-BBA-ADFAD, ELI280-CAA-ADHAX, ELI280-CEB-ACFBD, ELI280-AAB-ADCAD, ELI280-BDB-AACBD, ELI280-LDX-ADCBD, ELI280-AAB-ACCBX, ELI280-AAA-ABFBX, ELI280-DAB-ADCAD, ELI280-AFB-ABCBX, ELI280-DCB-AACAX, ELI280-DEB-ACFBD, ELI280-ADA-ACAAX, ELI280-AAB-ADFAD, ELI280-DCB-AAFBD, ELI280-CAA-AAAAX, ELI280-BDB-ACCAX, ELI280-CAA-AAFBX, ELI280-DDB-AAABX, ELI280-DCB-ACAAX, ELI280-CAA-ADFBD, ELI280-CBB-ACCBX, ELI280-DDB-AACBD, ELI280-DDB-ACCAX, ELI280-A, ELI280-DFA-ADCBX, ELI280-BCA-AAAAX, ELI280-DDB-AAAAX, ELI280-E, ELI280-C, ELI280-AAA-AAFBD, ELI280-DFC-ADFAD, BUR280-C1X, ELI280-D, ELI280-BAA-ACCBD, ELI280-AAA-AAAAX, BUR280-W1X, ELI280-B, ELI280-BCB-AAABX, ELI280-BCB-AACAX, BUR280-W1D, ELI280-F, ELI280-BCB-AAFAD, ELI280-BCB-BAFAX, BUR280-81D
- Recall number
- Z-0011-2022
- Recalling firm
- Welch Allyn Inc Mortara
- Product
- ELI, Burdick and McKesson brand 280 Resting Electrocardiographs (ELI 280, MLBUR280 and BUR280), Model Numbers: ELI280-DDB-ADAAX, ELI280-BDB-ACAAX, ELI280-CAA-AAFAD, ELI280-BDB-BDFAX, ELI280-BCB-AAAAX, ELI280-DDB-ACAAX, ELI280-DBA-BAFAX, ELI280-DCB-AAAAX, ELI280-DBA-AAFBD, ELI280-AAA-AAEBX, ELI280-BBA-ADFBD, ELI280-DDB-ACFAX, MLBUR280-81X, ELI280-CAA-ACEBX, ELI280-AAA-AAHBX, ELI280-DDB-BDFAX, ELI280-CAA-AAFBT, ELI280-DCD-ADFAD, ELI280-DDB-AACBX, ELI280-JXX-BDFAX, ELI280-CAA-ADCBX, ELI280-BCB-AACBD, ELI280-LDX-ADFBX, ELI280-LDX-ADCBX, ELI280-BDD-ADFAD, ELI280-BDB-AAFBD, ELI280-BDB-ACCAD, ELI280-BDB-ADFAD, ELI280-LDX-ADFBD, ELI280-DDB-AAFBD, ELI280-DCB-ACFAD, ELI280-AAA-BAFAF, MLBUR280-W1X, ELI280-DCB-AAABX, ELI280-AAA-ACAAX, ELI280-ADA-ABFBX, ELI280-BCB-AAFBD, ELI280-BCB-AAFBX, ELI280-CEA-ADFBX, ELI280-CDA-ADABX, ELI280-BDB-AAABX, ELI280-DDB-ACFBD, ELI280-DDB-BCFAX, ELI280-ADA-ADCAX, ELI280-BDB-AACBX, ELI280-BCB-ACAAX, BUR280-81X, ELI280-DCB-AAFBG, ELI280-BDB-AAFBT, ELI280-DCB-AAFAD, ELI280-CAA-ADFBX, ELI280-BDB-AAFBG, ELI280-BDB-AAFBX, ELI280-DCA-ACAAX, ELI280-CAB-ACFBX, ELI280-ADA-ACFAX, ELI280-CAA-AAFAT, ELI280-CAA-ACFBD, ELI280-CDA-ADCBX, ELI280-AAA-ADCBX, ELI280-CAA-AAFBD, ELI280-DCB-BAFBT, ELI280-BBA-AAAAX, ELI280-BBA-AAFAD, ELI280-DBA-AAFAD, ELI280-DCB-AAFBT, ELI280-DBA-ADFAX, ELI280-BBA-AAFBD, MLBUR280-C1X, ELI280-BBA-ADFAX, ELI280-DDD-ADFAD, ELI280-DBA-AAFAX, MLBUR280-W1D, ELI280-BCB-AACBX, ELI280-DCB-AACBD, ELI280-LDX-ADFBG, ELI280-DCB-AACBX, ELI280-DBA-ADFBD, ELI280-DDB-AAFBG, ELI280-DBA-AAABD, ELI280-DCB-AAFBX, ELI280-AAA-AAFBT, MLBUR280-C1D, ELI280-AAB-ADAAX, ELI280-DDB-AAFBT, ELI280-CAA-AACBX, ELI280-CEB-ACFBX, ELI280-CAA-ABFAX, ELI280-DDB-AAFBX, ELI280-DBA-AAAAX, ELI280-BCB-AAFBG, ELI280-BFA-ADCBX, ELI280-LDX-ADABX, MLBUR280-81D, ELI280-BBA-ADFAD, ELI280-CAA-ADHAX, ELI280-CEB-ACFBD, ELI280-AAB-ADCAD, ELI280-BDB-AACBD, ELI280-LDX-ADCBD, ELI280-AAB-ACCBX, ELI280-AAA-ABFBX, ELI280-DAB-ADCAD, ELI280-AFB-ABCBX, ELI280-DCB-AACAX, ELI280-DEB-ACFBD, ELI280-ADA-ACAAX, ELI280-AAB-ADFAD, ELI280-DCB-AAFBD, ELI280-CAA-AAAAX, ELI280-BDB-ACCAX, ELI280-CAA-AAFBX, ELI280-DDB-AAABX, ELI280-DCB-ACAAX, ELI280-CAA-ADFBD, ELI280-CBB-ACCBX, ELI280-DDB-AACBD, ELI280-DDB-ACCAX, ELI280-A, ELI280-DFA-ADCBX, ELI280-BCA-AAAAX, ELI280-DDB-AAAAX, ELI280-E, ELI280-C, ELI280-AAA-AAFBD, ELI280-DFC-ADFAD, BUR280-C1X, ELI280-D, ELI280-BAA-ACCBD, ELI280-AAA-AAAAX, BUR280-W1X, ELI280-B, ELI280-BCB-AAABX, ELI280-BCB-AACAX, BUR280-W1D, ELI280-F, ELI280-BCB-AAFAD, ELI280-BCB-BAFAX, BUR280-81D
- Status
- Open, Classified
- Initiated
- 2021-09-03
- Posted
- not on file
- Terminated
- not on file
- Reason
- The devices malfunction under specific operator workflows.
- Root cause
- Software design
- Action
- The firm notified its consignees on 9/3/2021 by letter via U PS or via USPS first class for those with P.O. Box mailing addresses. The letter notified customers of the issue, provided clinical mitigations and advised them not to utilize specific workflows until their software is updated.
- Quantity
- 14,630 units
- Distribution
- Distribution was nationwide, including Puerto Rico. Foreign distribution was made to Afghanistan, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Czechia, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Guatemala, Hongkong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Monaco, Morocco, Netherlands, New Zealand, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, The Netherlands, Trinidad and Tobago, Turkey, United Arab Emirates, and United Kingdom.
- Lot, serial or model codes
- Software versions 2.1.0-2.3.0, Serial numbers 116280503226 thru 121250000503 manufactured from 01-July-2016 thru 01-Jul-2021.
- 510(k) numbers
- K122073
- PMA numbers
- not on file
- Product code
- DPS
- Firm FEI
- 2183461
- Firm city
- Milwaukee
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28