Ortho-Clinical Diagnostics: VITROS Chemistry Products PHYT (Phenytoin) Slides, REF/Product Code 829 8671, IVD. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Recall number
Z-0007-2015
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS Chemistry Products PHYT (Phenytoin) Slides, REF/Product Code 829 8671, IVD. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Status
Terminated
Initiated
2014-08-20
Posted
2014-10-02
Terminated
2017-04-11
Reason
Ortho Clinical Diagnostics identified a potential for biased results to be generated when using VITROS PHYT Slides, Lot 2613-0150-4913. Some positively biased results were identified during testing of a different VITROS PHYT Slide lot; the affected slides were not released for distribution. Since Lot 2613-0150-4913 was manufactured in the same event, there is the potential for slides from this lot to also be affected.
Root cause
Software Manufacturing/Software Deployment
Action
On 8/20/14, an URGENT PRODUCT CORRECTION NOTIFICATION Letter (CL2014-222, dated 8/20/14) was sent via FedEx overnight courier and/or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to US consignees (OCD Direct, Drop-Ship and US Federal Government) to inform them of the issue and advise them to discontinue use of the affected lot(s) and to inform their Laboratory Medical Director if they suspect that a previously reported result may have been affected. One foreign affiliate was notified on 8/20/14. One distributor was notified on 8/20/14. The customers were advised to: - Immediately discontinue using and discard VITROS Chemistry Products PHYT slide Lot 2613-0150-4913. Review VITROS PHYT results obtained using Lot 2613-0150-4913 and discuss any concerns they may have regarding previously reported results with their Laboratory Medical Director or with the requesting physician. Complete the attached Confirmation of Receipt form no later than 8/28/14. Forward the information in this notification if they have distributed this product outside of their facility facility.
Quantity
Domestic: 339 sales units; Foreign: 10 sales units
Distribution
Distributed USA (nationwide) and in Singapore.
Lot, serial or model codes
Lot Number 2613-0150-4913, exp. 15-Nov-2014
510(k) numbers
["K941142"]
PMA numbers
not on file
Product code
DIP
Firm FEI
1000305840
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register