Skip to the content

FDA Medical Device Recalls (openFDA)

Sedecal USA, Inc: Sedecal SP-HF 4.0 Portable X-Ray System; Manufactured by SEDECAL, C/Pelya 9 Pol. Ind. Rio De Janeiro, 28110 Algete (Madrid) Spain

Recall number
Z-0001-04
Recalling firm
Sedecal USA, Inc
Product
Sedecal SP-HF 4.0 Portable X-Ray System; Manufactured by SEDECAL, C/Pelya 9 Pol. Ind. Rio De Janeiro, 28110 Algete (Madrid) Spain
Status
Terminated
Initiated
2003-10-27
Posted
2003-10-29
Terminated
2004-03-24
Reason
A minimum source-skin-distance of less than 30 cm and not identifying the tube manufacturer on the tube housing label resulted in the SP-HF-4.0 Portable Systems not complying with the U.S. Federal performance standard.
Root cause
Other
Action
An 'Important Safety Notice' dated 10/27/03 was sent to the one direct account on the same date, informing them of the non-compliances with the performance standard and requesting them to call Sedecal at 800-920-9525 to arrange for the modification of the x-ray unit at no cost to the user.
Quantity
1 unit
Distribution
Vermont
Lot, serial or model codes
serial #PK10058
510(k) numbers
["K020436"]
PMA numbers
not on file
Product code
IZL--
Firm FEI
1000139935
Firm city
Arlington Heights
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register