Skip to the content

FDA Medical Device Recalls (openFDA)

Cardinal Health 200, LLC: (a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888266114, Sterile; (b) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) x 48 in. (122 cm), Model #8888266122, Sterile; (c) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 16 Fr/Ch (5.3 mm) x 48 in. (122 cm), Model #8888266130, Sterile; and (d) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 18 Fr/Ch (6.0 mm) x 48 in. (122 cm), Model #8888266148, Sterile.

Recall number
Z-3302-2024
Recalling firm
Cardinal Health 200, LLC
Product
(a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888266114, Sterile; (b) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) x 48 in. (122 cm), Model #8888266122, Sterile; (c) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 16 Fr/Ch (5.3 mm) x 48 in. (122 cm), Model #8888266130, Sterile; and (d) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 18 Fr/Ch (6.0 mm) x 48 in. (122 cm), Model #8888266148, Sterile.
Status
Open, Classified
Initiated
2024-07-24
Posted
2024-09-26
Terminated
not on file
Reason
The firm received reports of breakage in the Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.
Root cause
Labeling Change Control
Action
The recalling firm issued letters dated 7/22/2024 via overnight mail on 7/24/2024. The letter informed the consignee the product labeling has been updated via electronic instructions for use for the affected product due to reported incidences of improper use of the device which can lead to breakage of the Salem Sump Anti-Reflux Valve (ARV) and an increased risk to patients. To mitigate the risk of improper use, the instructions have been revised. The risk to health and actions required were included in the letter. The actions included: (1) Review your inventory for the affected product codes; (2) Communicate the change of the use instructions with all personnel that utilize the devices; (3) Post a copy of Attachment A (the revised IFU) and the recall notification in the storeroom and clinical areas; (4) Notify any customers to whom you may have distributed/forwarded affected product or to whom you intend to distribute/forward product) regarding this medical device product correction and share a copy of the notice; (5) Return the enclosed acknowledgment form via FAX or email.
Quantity
24,761,100 tubes total
Distribution
Distribution was nationwide, including Puerto Rico and Guam. There was also government/military distribution. Foreign distribution was made to Chile.
Lot, serial or model codes
All lot numbers. UDI-DI numbers: (a) 8888266114, UDI-DI 10192253012781; (b) 8888266122, UDI-DI 10192253012804; (c) 8888266130, UDI-DI 10192253012828; and (d) 8888266148, UDI-DI 10192253012842.
510(k) numbers
["K880850"]
PMA numbers
not on file
Product code
FEG
Firm FEI
3001236905
Firm city
Waukegan
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register