Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Comprehensive Reverse Humeral Fracture Inserter REF 406993 Comprehensive Instrumentation Reverse Humeral Inserter QTY 1. Instrument that is attached to the trial prosthesis for placement in the prepared bone canal to check for correct sizing, version and humeral length. Also used to insert the humeral implant into the canal.

Recall number
Z-3289-2011
Recalling firm
Biomet, Inc.
Product
Comprehensive Reverse Humeral Fracture Inserter REF 406993 Comprehensive Instrumentation Reverse Humeral Inserter QTY 1. Instrument that is attached to the trial prosthesis for placement in the prepared bone canal to check for correct sizing, version and humeral length. Also used to insert the humeral implant into the canal.
Status
Terminated
Initiated
2011-08-03
Posted
2011-09-30
Terminated
2011-10-26
Reason
Recall initiated in response to a Biomet Canada report, which noted that while testing the instruments, the pin used to secure the implant on the device did not engage on the effected lots and were distributed.
Root cause
Nonconforming Material/Component
Action
On 8/3/11 Biomet Orthopedics sent an "URGENT MEDICAL DEVICE RECALL NOTICE" letter to all affected customers. The letter included: reason for the recall, risks associated with use of the device, and alternate instructions for use of the recalled device. Customers are instructed to locate the affected products, discontinue use, return the devices, and a fax a copy of the response form. For information or questions on this recall call Biomet at (574) 372-1570.
Quantity
6 units (internationally)
Distribution
International Distribution Only: Canada, Australia, and Columbia.
Lot, serial or model codes
Part numbers: 406993, Lot numbers:176070 and 657550; and 406993-00, Lot number : 203119.
510(k) numbers
["K023063"]
PMA numbers
not on file
Product code
LXH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register