Skip to the content

FDA Medical Device Recalls (openFDA)

Rayner Intraocular Lenses Limited: C-flex 570C +19.0D Intraocular Lens Product Usage: C-flex IOLs are designed to be surgically implanted into the capsular bag of the human eye as a replacement for the crystalline lens following phacoemulsification

Recall number
Z-3282-2018
Recalling firm
Rayner Intraocular Lenses Limited
Product
C-flex 570C +19.0D Intraocular Lens Product Usage: C-flex IOLs are designed to be surgically implanted into the capsular bag of the human eye as a replacement for the crystalline lens following phacoemulsification
Status
Terminated
Initiated
2018-08-09
Posted
not on file
Terminated
2020-11-24
Reason
Firm become aware of reports of post-operative refractive errors following implantation of lenses.
Root cause
Error in labeling
Action
On August 31 the firm sent letters to all customers (health care facilities) that have received lenses in this batch and instructed to quarantine any remaining unused lenses from the C-flex 570C +19.0D batch 017100825. Replacement, reimbursement or substitution is offered (as appropriate).
Quantity
49
Distribution
US in the states of MO
Lot, serial or model codes
Batch 017100825 Lens #'s: 01710082501, 01710082502, 01710082503, 01710082504, 01710082505, 01710082506, 01710082507, 01710082508, 01710082509, 01710082510, 01710082511, 01710082512, 01710082513, 01710082514, 01710082515, 01710082516, 01710082517, 01710082518, 01710082519, 01710082520,  01710082521, 01710082522, 01710082523, 01710082524, 01710082525, 01710082526, 01710082527, 01710082528, 01710082529, 01710082530, 01710082531, 01710082532, 01710082533, 01710082534, 01710082535, 01710082536, 01710082537, 01710082538, 01710082539, 01710082540,  01710082541, 01710082542, 01710082543, 01710082544, 01710082545, 01710082546, 01710082547, 01710082548, and 01710082549.
510(k) numbers
not on file
PMA numbers
["P060011"]
Product code
HQL
Firm FEI
3012304651
Firm city
Worthing
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register