FDA Medical Device Recalls (openFDA)
Ortho-Clinical Diagnostics: VITROS Sodium Slides, Catalog # 837 9034
- Recall number
- Z-3280-2018
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- VITROS Sodium Slides, Catalog # 837 9034
- Status
- Terminated
- Initiated
- 2018-05-05
- Posted
- not on file
- Terminated
- 2021-02-19
- Reason
- Potential for positively biased results
- Root cause
- Nonconforming Material/Component
- Action
- On 05 May 2018, Customer letter (Ref. CL2018-123) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all VITROS customers, who have a analyzer capable of running VITROS Na+ slides, to inform them of the issue and to request that they immediately discontinue use of this VITROS Na+ Slides, GENs 8 and 13. Continued use of VITROS Na+ Slides, GENs14,16,17& 18 is acceptable provided the conditions detailed in the communication are met. Foreign affiliates were initially informed by email on 05 May 2018 of the issue and instructed to notify their consignees of the affected product with this issue and the requirement to immediately discontinue use of the VITROS Na+ slides GENs 8, 13, 14, 16, 17 and 18. Due to the amount of time required to ship replacement product to some worldwide locations, a revised communication to the foreign affiliates occurred on 08 June 2018. Due to the health risk associated with potentially leaving the laboratory without a Na+ method, the revised letter instructs customers that they may continue to use GENs 14, 16, 17 and 18 until replacement product is received.
- Quantity
- not on file
- Distribution
- US Nationwide, Bermuda, Austria, Brazil, Canada, Colombia, China, India, Japan, Mexico, Singapore, UK, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium and the Netherlands
- Lot, serial or model codes
- UDI - 10758750004812 GENs- 8, 13, 14, 16, 17 & 18 Expiry Dates: 01-Nov-2018 through 01-Nov-2019
- 510(k) numbers
- ["K924471"]
- PMA numbers
- not on file
- Product code
- JGS
- Firm FEI
- 1000305840
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28