Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: PRO+ MRS SURFACE FOR VERSACARE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intake and output valves to provide weight-based pressure redistribution whenever a patient moves or is repositioned, Model/Catalog Number: P3255A08

Recall number
Z-3262-2026
Recalling firm
Baxter Healthcare Corporation
Product
PRO+ MRS SURFACE FOR VERSACARE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intake and output valves to provide weight-based pressure redistribution whenever a patient moves or is repositioned, Model/Catalog Number: P3255A08
Status
Open, Classified
Initiated
2026-08-18
Posted
2026-09-24
Terminated
not on file
Reason
There is an error in the product description on the Unique Device Identification (UDI) label. These discrepancies in labeling do not align with the approved product or product specifications and may lead to incorrect product identification or selection, potentially resulting in use with an incompatible bed frame.
Root cause
Process control
Action
An "Urgent Medical Device Correction notification" letter dated 8/18/26 was sent to customers. Actions to be Taken by Customers 1.Locate the affected product at your facility and confirm each Product Code. Once product has been identified, use Hillrom pro+ Surface Instructions for Use (209196) (IFU) to assess the fit of the surfaces relative to the frame and siderails to confirm the surface is installed properly. If the surface does not meet the requirements set in the IFU please discontinue the use of this surface. 2. A Baxter representative will contact your facility to replace the incorrect label when it becomes available. 3. If you received this communication directly from Baxter, please acknowledge receipt of this notification by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com even if you do not have any remaining inventory. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from Accenture LLP on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 5. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction. For general questions regarding this communication, please contact Baxter Global Technical Service at 800-445-3720 Option 2 between 8:00 am and 6:00 pm Eastern Time Monday through Friday.
Quantity
169 units
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Lot, serial or model codes
Serial Number: A171BP3769 - B057BP6765 UDI-DI Number: 00887761007543
510(k) numbers
not on file
PMA numbers
not on file
Product code
IOQ
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register