FDA Medical Device Recalls (openFDA)
Angiodynamics, Inc.: ANGIODYNAMICS NANOKNIFE IRE GENERATOR, Catalog/REF No. 20300101, ANGIODYNAMICS, 603 Queensbury Ave., Queensbury, NY 12804 Intended to remove tissue and control bleeding by the use of high- frequency electrical current (surgical ablation of soft tissue).
- Recall number
- Z-3258-2011
- Recalling firm
- Angiodynamics, Inc.
- Product
- ANGIODYNAMICS NANOKNIFE IRE GENERATOR, Catalog/REF No. 20300101, ANGIODYNAMICS, 603 Queensbury Ave., Queensbury, NY 12804 Intended to remove tissue and control bleeding by the use of high- frequency electrical current (surgical ablation of soft tissue).
- Status
- Terminated
- Initiated
- 2011-04-04
- Posted
- 2011-09-23
- Terminated
- 2016-01-29
- Reason
- Some units of Nanoknife have the potential to deliver a shortened pulse width without indicating an error. The affected units contain a "GOMe" circuit board which may cause the generator to drift from the calibration points set at the time of manufacture.
- Root cause
- Component design/selection
- Action
- AngioDynamics, Inc. sent a "NOTIFICATION OF MEDICAL DEVICE CORRECTION" letter dated April 4, 2011 to all affected customers. The letter describes the product, problem, and action to be taken. The letter instructs customers to review the measurement graphs (displayed on generator) after each delivery of energy, and to contact AngioDynamics Customer Service at 1-800-772-6446 if shortened pulses are detected. Contact your local sales representative or call Customer Service for any additional concerns or questions regarding this notice.
- Quantity
- 21 units
- Distribution
- Worldwide Distribution--USA (nationwide) and the countries of Australia, Germany, Israel, Italy, and Spain,
- Lot, serial or model codes
- Batch/Serial Numbers: 00120508, 00090508, 00880909, 00371008, 00670709, 00850909, 00890909, 00780709, 00220708, 00300708, 00951009, 00900909, 00800709, 00260708, 00650709, 00690709, 00920909, 00930909, 00770709, 00130508, 01001009.
- 510(k) numbers
- ["K080376"]
- PMA numbers
- not on file
- Product code
- OAB
- Firm FEI
- 1319211
- Firm city
- Queensbury
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28