FDA Medical Device Recalls (openFDA)
Boston Scientific Corporation: ESSENTIO Pacemaker
- Recall number
- Z-3255-2018
- Recalling firm
- Boston Scientific Corporation
- Product
- ESSENTIO Pacemaker
- Status
- Terminated
- Initiated
- 2018-09-10
- Posted
- not on file
- Terminated
- 2019-11-04
- Reason
- Boston Scientific has identified a subset of active pacemaker and CRT-Ps systems with an elevated potential for early pacemaker replacement due to hydrogen-induced accelerated battery depletion.
- Root cause
- Component design/selection
- Action
- On September 2018 Boston Scientific sent letters to their customers with the following information: Dear Doctor, Boston Scientific has identified a subset of approximately 2,900 active ACCOLADE", PROPONENT", and ESSENTIO" pacemakers and VISIONIST" and VALITUDE" cardiac resynchronization therapy pacemakers (CRT-Ps) with an elevated potential for early pacemaker replacement due to hydrogen-induced accelerated battery depletion. There have been no reports of injury related to this behavior. You are receiving this letter because you or your center has implanted or follow one or more patients with a pacemaker in this subset. This device behavior can be identified via the regular pacemaker follow-up process, either in-clinic or through the LATITUDE" NXT Remote Patient Management System (LATITUDE). Therefore, a follow-up interval of no more than six (6) months is recommended and is consistent with existing international societal guidelines1. If accelerated depletion is suspected, Boston Scientific recommends consultation with Technical Services to review the pacemaker diagnostic data available from either a recent data upload in LATITUDE or a Save to Disk to confirm accelerated depletion and determine an appropriate timeframe for pacemaker replacement. Description The pacemaker manual describes how increases in pacing requirements or changes in programmed parameters may result in an expected longevity reduction and are considered normal battery depletion. However, hydrogen exposure within the pacemakers circuitry may compromise the electrical performance of low voltage capacitors causing current leakage and a moderate acceleration in the rate of battery depletion. Because this accelerated depletion does not occur rapidly, a follow-up interval of no more than six months is recommended. Boston Scientific has determined a liner component to be the source of hydrogen and identified a subset of previously distributed pacemakers that have an elevated potential for exh
- Quantity
- not on file
- Distribution
- AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, W A, WI, WV, and WY
- Lot, serial or model codes
- ESSENTIO SR model L100, ESSENTIO DR model L101, ESSENTIO MRI SR model L110, ESSENTIO MRI DR model L111, ESSENTIO EL DR model L121, and ESSENTIO MRI EL DR model L131
- 510(k) numbers
- not on file
- PMA numbers
- ["P150012"]
- Product code
- LWS
- Firm FEI
- 2124215
- Firm city
- Saint Paul
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28