Philips North America: IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)

Recall number
Z-3249-2024
Recalling firm
Philips North America
Product
IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)
Status
Open, Classified
Initiated
2024-08-30
Posted
2024-09-25
Terminated
not on file
Reason
During a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the devices electromagnetic immunity and emission.
Root cause
Nonconforming Material/Component
Action
An URGENT Medical Device Correction notification letter dated 8/30/24 was sent to customers. Actions that should be taken by the customer / user to prevent risks for patients or users " Please check if the equipotential connector is broken off (refer to Figure 1). If yes, remove the device from use. " Pass this notice to all those who need to be aware within your organization or to any organization where affected devices have been potentially transferred. " Customers should complete the URGENT Medical Device Response Form either via the QR code below or at the end of the notification to submit both their acknowledgment of this recall and confirm understanding of actions to be taken. If hard copy of Response Form is completed in lieu of digital version (via QR code) please return to Philips at recall.responsephilips.com. You may also fax it to 877 499 7223. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will replace the power supply. If you need any further information, please contact your local Philips representative, or call 1-800-722-9377. Distributors: Philips is requesting customers to return a Response Form to acknowledge receipt and understanding of the URGENT Medical Device Correction and confirm that the information from this Letter has been properly distributed to all users that handle the affected product. Together with this letter we are providing a list of affected products that Philips has sold to your organization. As distributor of the affected products, we kindly request that you: o Add in the Response Form attached your contact information. o Send the attached URGENT Medical Device Correction to each customer to whom you have distributed any affected product as soon as possible and no later than three days, together with the Reply Card. o Perform a good faith effort to get the Reply Form by following up with the customer with a minimum of t
Quantity
545 devices
Distribution
US Nationwide. Global Distribution.
Lot, serial or model codes
IntelliVue Patient Monitor MX400, 866060, UDI-DI: 00884838038752; IntelliVue Patient Monitor MX430, 866061, UDI-DI: 00884838057562; IntelliVue Patient Monitor MX450, 866062, UDI-DI: 00884838038769; IntelliVue Patient Monitor MX500, 866064, UDI-DI: 00884838038776; IntelliVue Patient Monitor MX550, 866066, UDI-DI: 00884838038783; Only MX400-550 devices shipped after 26-April-2024 are affected. Please refer to the manufacturing date on the back of your monitor.
510(k) numbers
["K130849","K131872"]
PMA numbers
not on file
Product code
MHX
Firm FEI
3006648320
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register