Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: Vanguard 360 Revision Knee System 63x 10mm Universal Tibial Block Augment, Model Number 185221 Intended for use in Total Knee Arthroplasty

Recall number
Z-3249-2018
Recalling firm
Zimmer Biomet, Inc.
Product
Vanguard 360 Revision Knee System 63x 10mm Universal Tibial Block Augment, Model Number 185221 Intended for use in Total Knee Arthroplasty
Status
Terminated
Initiated
2018-08-02
Posted
not on file
Terminated
2019-04-25
Reason
The implant only has one bolt in the sterile package but it should have had two.
Root cause
Under Investigation by firm
Action
The firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE RECALL" notification letters dated 8/2/18 to customers on 8/2/18. The letters described the product, problem and actions to be taken. The letters instruct customers to perform the following actions: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Distributors: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to CorporateQuality.PostMarket@zimmerbiomet.com within three (3) days. b. For each return, send a copy of Attachment 1 to CorporateQuality.PostMarket@zimmerbiomet.com. c. Include a hardcopy of Attachment 1 in each carton of your retur
Quantity
16
Distribution
Worldwide Distribution: US (nationwide) to the following US states: AZ, KS, NM, NV, and UT; and to countries of: Australia and Netherlands.
Lot, serial or model codes
Lot 320390, UDI (01) 0088030448157 (17) 230129 (10) 320390
510(k) numbers
["K093293"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register