FDA Medical Device Recalls (openFDA)
Zimmer Biomet, Inc.: Zimmer Natural Nail Tibial Nail - Yellow 8.3mm, Model Number 47249534008 Product Usage - The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone. Indications for the Tibial nails include the following in the tibia: Compound and simple shaft fractures Proximal, metaphyseal and distal shaft fractures Segmental fractures Comminuted fractures Fractures involving osteopenic and osteoporotic bone Pathological fractures Fractures with bone loss Pseudoarthrosis, non-union, mal-union and delayed union Periprosthetic fractures Surgically created defects such as osteotomies. Product Usage - The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone. Indications for the Tibial nails include the following in the tibia: Compound and simple shaft fractures Proximal, metaphyseal and distal shaft fractures Segmental fractures Comminuted fractures Fractures involving osteopenic and osteoporotic bone Pathological fractures Fractures with bone loss Pseudoarthrosis, non-union, mal-union and delayed union Periprosthetic fractures Surgically created defects such as osteotomies.
- Recall number
- Z-3246-2018
- Recalling firm
- Zimmer Biomet, Inc.
- Product
- Zimmer Natural Nail Tibial Nail - Yellow 8.3mm, Model Number 47249534008 Product Usage - The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone. Indications for the Tibial nails include the following in the tibia: Compound and simple shaft fractures Proximal, metaphyseal and distal shaft fractures Segmental fractures Comminuted fractures Fractures involving osteopenic and osteoporotic bone Pathological fractures Fractures with bone loss Pseudoarthrosis, non-union, mal-union and delayed union Periprosthetic fractures Surgically created defects such as osteotomies. Product Usage - The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone. Indications for the Tibial nails include the following in the tibia: Compound and simple shaft fractures Proximal, metaphyseal and distal shaft fractures Segmental fractures Comminuted fractures Fractures involving osteopenic and osteoporotic bone Pathological fractures Fractures with bone loss Pseudoarthrosis, non-union, mal-union and delayed union Periprosthetic fractures Surgically created defects such as osteotomies.
- Status
- Terminated
- Initiated
- 2018-07-30
- Posted
- not on file
- Terminated
- 2019-10-24
- Reason
- The item is labeled as an 8.3mm, but it is actually a 12 mm.
- Root cause
- Mixed-up of materials/components
- Action
- On 7/30/18, Urgent Medical Device Recall notification letters were distributed to customers. Distributors were instructed to do the following: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals and surgeons within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to CorporateQuality.PostMarket@zimmerbiomet.com within three (3) days. b. Request a Return Authorization Number through zimmerbiometintlirarequests@zimmerbiomet.com for international returns c. For each return, send a copy of Attachment 1 to CorporateQuality.PostMarket@zimmerbiomet.com d. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. e. Mark RECALL on the outside of the returned cartons. 4. Retain a copy of your Inventory Return Certification and product return forms for your records in the event of a compliance audit of your facility. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Risk Managers and Surgeons were asked to do the following: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement an
- Quantity
- 1 unit
- Distribution
- The product was not distributed in the United States. The product was distributed in the Netherlands.
- Lot, serial or model codes
- UDI: (01)00889024097551(17)190228(10)61169228R, Lot 61169228R
- 510(k) numbers
- ["K082770"]
- PMA numbers
- not on file
- Product code
- HSB
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28