FDA Medical Device Recalls (openFDA)
Philips Electronics North America Corporation: Cesar-DMCP-Visub(HM2000/3000), System Code 72239 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
- Recall number
- Z-3238-2018
- Recalling firm
- Philips Electronics North America Corporation
- Product
- Cesar-DMCP-Visub(HM2000/3000), System Code 72239 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
- Status
- Terminated
- Initiated
- 2018-07-16
- Posted
- 2018-09-21
- Terminated
- 2022-11-18
- Reason
- There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
- Root cause
- Device Design
- Action
- The firm, BU IGT System, sent an "Urgent -Medical Device Correction" notification letter dated 07/16/2018 to customers. The letter describes the product, problem and actions to be taken. The letter instructs customers to do the following: In order to reduce the risk for patients, users and bystanders if this problem would reoccur, we recommend the following actions until the correction has been implemented. -Avoid unnecessary movements of the Monitor Ceiling Suspension. -For those movements that are necessary, avoid that the user, patient or bystander are in close proximity to the monitor. -When moving the Monitor Ceiling Suspension, ensure that no body partsof the staff or patient are underneath the monitor. -Do not move the monitor above the patient. Please ensure that all staff with access to the affected systems are informed of the content of this Medical Device Correction. All affected products will be corrected by means of a Field Change Order (FCO) free of charge. This FCO will be available mid-August. You will be contacted by our local Philips representative to schedule this corrective action. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722- 9377, 8am to 8pm (EST) Monday thru Friday.
- Quantity
- 5115 total
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and countries of: Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, British Virgin Islands, Brazil, Brunei Darussalam, Blugaria, Cambodia, Canada, Chile, China, Columbia, Croatia, Cyprus, Czech Republic, Denmark, Dominic Republic, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea South, Republic of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Reunion, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, UAE, UK, Uzbekistan, Venezuela, Vietnam, and Yemen.
- Lot, serial or model codes
- All systems with an actuator for the FlexVision Monitor Ceiling Suspension in the period 2003 to May 2011 are affected
- 510(k) numbers
- ["K925302"]
- PMA numbers
- not on file
- Product code
- IZF
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28