Skip to the content

FDA Medical Device Recalls (openFDA)

Bomimed: BOMImed Disposable Fiber Optic Medium Laryngoscope Handle. Each handle is packaged in an individual peel pouch, 20 handles per box. The rigid laryngoscope handle is a device used along with a blade to examine and visualize a patient's airway and aid placement of a tracheal tube.

Recall number
Z-3238-2017
Recalling firm
Bomimed
Product
BOMImed Disposable Fiber Optic Medium Laryngoscope Handle. Each handle is packaged in an individual peel pouch, 20 handles per box. The rigid laryngoscope handle is a device used along with a blade to examine and visualize a patient's airway and aid placement of a tracheal tube.
Status
Terminated
Initiated
2017-07-13
Posted
not on file
Terminated
2021-04-01
Reason
Potential for intermittent lighting (flickering) during use.
Root cause
Device Design
Action
A letter was sent to all the consignees via mail dated July 13, 2017 informing of the recall due to Length variation of the battery contact inside the handle due to manufacturing/assembly variation may potentially cause an intermittent electrical connection resulting in flickering light. Complete the bottom section and return by email to regulatory@bomimed.com to confirm this notice has been received and understood. Check functionality of the light for each handle by engaging a blade and applying pressure to simulate use. Complete this task prior to use. Contact BOMImed for replacements if intermittent lighting is present. Replacement product orders will be organized in a manner to assist in covering all affected customers needs. Contac: Trina Friesen Quality Assurance & Regulatory Manager Tel: 800-667-6276 ext.234 Fax: 877-435-6984 regulatory@bomimed.com
Quantity
680 (US) 2,220 (OUS)
Distribution
IL. PA. and WA CANADA
Lot, serial or model codes
Part number: OL-334L6 Lot#: 2216, 3016, 4016 and 4816
510(k) numbers
not on file
PMA numbers
not on file
Product code
CCW
Firm FEI
3003213883
Firm city
Winnipeg
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register