FDA Medical Device Recalls (openFDA)

Radiometer America Inc: AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality control (QC) ampoules are specifically designed for Radiometer America s blood gas analyzers

Recall number
Z-3237-2017
Recalling firm
Radiometer America Inc
Product
AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality control (QC) ampoules are specifically designed for Radiometer America s blood gas analyzers
Status
Terminated
Initiated
2017-09-07
Posted
not on file
Terminated
2019-08-01
Reason
Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92 5.85 kPa or 36.9 46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)
Root cause
Under Investigation by firm
Action
Radiometer sent a Safety Notice dated September 7, 2017, to all affected customers. The firm initiated their recall on 09/07/2017 by email. The letter requested the following: 1. Please check your inventory and analyzers for any boxes and ampoules of above mentioned QC. 2. Discard all Boxes and ampoules of this lot. 3. Complete the response form and return it to the number on the form. Please note: If you are not the end-user of the affected product, please ensure that this letter is distributed to the final end-user. Customers with questions were advised to contact their Radiometer representative.
Quantity
1669 Boxes
Distribution
Worldwide Distribution - US (nationwide) Internationally to Canada, Netherlands, China, Czech Republic, Denmark, Germany, Spain France, Hungary, India, Denmark, Tokyo, South Korea, United Kingdom, Australia, Poland, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, and Finland.
Lot, serial or model codes
S7745 Lot R0698
510(k) numbers
["K980135"]
PMA numbers
not on file
Product code
JJS
Firm FEI
3010420736
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register