Skip to the content

FDA Medical Device Recalls (openFDA)

NOVADAQ TECHNOLOGIES INC.: NOVADAQ SPY-PHI PAQ Intended for use with the SPY-PHI System, Product Code HH9006, containing 6 Single-Procedure Kits (Product Code HH9001), each containing the sterile drape (HH2000), as well as single-use drug components, Manufactured for and Distributed by Novadaq Technologies Inc., Burnaby, BC, Canada, Assembled by Carestream Health, Inc., Rochester, NY. The SPY-PHI Drape is intended to cover the SPY-PHI System to maintain a sterile field, and as an aid in the clean-up of equipment after surgery or other medical procedures. The product is not intended to be used as patient drapes and they have no patient contact.

Recall number
Z-3235-2017
Recalling firm
NOVADAQ TECHNOLOGIES INC.
Product
NOVADAQ SPY-PHI PAQ Intended for use with the SPY-PHI System, Product Code HH9006, containing 6 Single-Procedure Kits (Product Code HH9001), each containing the sterile drape (HH2000), as well as single-use drug components, Manufactured for and Distributed by Novadaq Technologies Inc., Burnaby, BC, Canada, Assembled by Carestream Health, Inc., Rochester, NY. The SPY-PHI Drape is intended to cover the SPY-PHI System to maintain a sterile field, and as an aid in the clean-up of equipment after surgery or other medical procedures. The product is not intended to be used as patient drapes and they have no patient contact.
Status
Terminated
Initiated
2017-06-27
Posted
not on file
Terminated
2017-12-11
Reason
The sterile drape material may contain a pinhole compromising its integrity resulting in a breach of the sterile barrier.
Root cause
Under Investigation by firm
Action
The recalling firm, Novadaq, issued an "URGENT DEVICE RECALL" letter dated 6/15/2017 via delivery by firm sales representatives or courier service (such as FedEx or UPS) on 6/27/2017 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to return unused product to Novadaq representative and complete and return the Acknowledgement of Receipt form by email to Novadaq Quality Department at quality@novadaq.com within 24 hours of receipt. Any questions or concerns may also be sent the listed email address. If you have any questions, please do not hesitate to contact the Director of Regulatory Affairs at 1.905.629.3822 ext 205 or email: jpendlebury@novadaq.com.
Quantity
16/6-drape kits
Distribution
Worldwide Distribution. US (nationwide) to states of: CA, CO, ID, IL, NY, OR, and VA; and countries of: Canada, Belgium, and Lithuania.
Lot, serial or model codes
PAQ lot #1704B containing drape lot #1703251; PAQ lot #1704F containing drape lot #1703291; and PAQ lot #17046 containing drape lot #1703251.
510(k) numbers
["K041501"]
PMA numbers
not on file
Product code
MMP
Firm FEI
3012345110
Firm city
Burnaby
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register