Skip to the content

FDA Medical Device Recalls (openFDA)

Fujifilm Medical Systems U.S.A., Inc.: FUJINON ED-530XT. This device is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Recall number
Z-3225-2017
Recalling firm
Fujifilm Medical Systems U.S.A., Inc.
Product
FUJINON ED-530XT. This device is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.
Status
Terminated
Initiated
2017-07-21
Posted
not on file
Terminated
2018-07-12
Reason
An update to the design and labeling was implemented to help reduce patient risk associated with inadequate reprocessing of the device. The action includes replacement of the forceps elevator mechanism, the O-ring seal, the distal end cap and issuance of a new Operation Manual.
Root cause
Device Design
Action
On July 21, 2017, Fujifilm Medical Systems U.S.A., Inc. Endoscopy Division distributed Urgent Medical Device Correction and Removal notices & Field Action Verification Forms to their U.S. Consignees by USPS certified mail. Fujifilm Medical Systems will contact all customers that have the duodenoscope by October 2017, with details for returning the device for replacement parts. This field action allows continued use of the current device until you are contacted by a sales representative for replacement parts as early as October 2017. While the remediation of the scopes will take approximately two weeks to complete, Fujifilm Medical Systems U.S.A. Inc. Endoscopy Division will provide interim duodenoscopes on loan for continuity of care. Once received, the recalling firm will replace the forceps elevator mechanism with O-ring and the distal end cap. An updated manual has been generated to include new information/guidance. Customers are encouraged to complete and return the Field Action Verification Form via email to fnonmedicalqa@fujifilm.com or via fax to (973) 872-4723. Customers with questions can call Larry Picciano, Sr. Director Quality and Regulatory Affairs at (973) 686-2479.
Quantity
362 units
Distribution
Nationwide
Lot, serial or model codes
All lots/serial numbers
510(k) numbers
["K152257"]
PMA numbers
not on file
Product code
FDT
Firm FEI
2431293
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register