Becton Dickinson & Co.: BD MAX RNA Extraction Kit RNA-3, Catalog Number 437506. Kit contains 24/2D barcode sample preparation reagent tubes. The BD MAX RNA Extraction Kit is used with the BD MAX System to extract RNA from viruses which maybe present in clinical specimens. Purified RNA obtained with the BD MAX System or another commercially available system for reverse transcription followed by nucleic acid amplification and detection. The BD MAX RNA Extraction Kit has not been validated for use with any specific analytical test method. The BD MAX DNA extraction kit is used with the BD MAX System to extract DNA from Gram-negative bacteria and viruses, which may be present in clinical specimens.

Recall number
Z-3219-2011
Recalling firm
Becton Dickinson & Co.
Product
BD MAX RNA Extraction Kit RNA-3, Catalog Number 437506. Kit contains 24/2D barcode sample preparation reagent tubes. The BD MAX RNA Extraction Kit is used with the BD MAX System to extract RNA from viruses which maybe present in clinical specimens. Purified RNA obtained with the BD MAX System or another commercially available system for reverse transcription followed by nucleic acid amplification and detection. The BD MAX RNA Extraction Kit has not been validated for use with any specific analytical test method. The BD MAX DNA extraction kit is used with the BD MAX System to extract DNA from Gram-negative bacteria and viruses, which may be present in clinical specimens.
Status
Terminated
Initiated
2011-05-24
Posted
2011-09-16
Terminated
2011-10-12
Reason
Sample results maybe switched between time it was scanned with handheld barcode reader and time it was placed in the sample rack, associating sample results to wrong patient.
Root cause
Nonconforming Material/Component
Action
BD Diagnostic Systems sent an "URGENT PRODUCT RECALL" letter dated May 2011, to all affected customers. The letter identified the products, the problem, and the action to be taken by the customers. The letter advised customers to discontinue use of the Sample Preparation Reagent tubes in the referenced kits and discard any remaining products. Customers were requested to complete an attached reply form to acknowledge their receipt of the notification and fax it to (410) 316-4258. For further assistance regarding replacements contact BD Customer Service Dept. 1-800-675-0908. For all other inquiries, contact BD Technical Services Dept. 1-800-638-8663.
Quantity
138 kits
Distribution
Worldwide Distribution-USA (nationwide) including the states of CA, KY, MD, MI, NC, NY, TX, and WI and the country Belgium.
Lot, serial or model codes
Lot numbers 103551, 110071, 110381, 110382, 110383, 110841, 110842, 110901, 110902
510(k) numbers
not on file
PMA numbers
not on file
Product code
NQX
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register