Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Neurosurgery: Medtronic Clip Gun Magazines are packaged in individually sealed pouches and distributed in boxes containing 5 individually pouched Clip Gun Magazine. REF CM8902, Medtronic Clip Gun Magazine, Rx only, CE 0344, STERILE EO, Rx only, Single Use

Recall number
Z-3212-2018
Recalling firm
Medtronic Neurosurgery
Product
Medtronic Clip Gun Magazines are packaged in individually sealed pouches and distributed in boxes containing 5 individually pouched Clip Gun Magazine. REF CM8902, Medtronic Clip Gun Magazine, Rx only, CE 0344, STERILE EO, Rx only, Single Use
Status
Terminated
Initiated
2018-05-03
Posted
not on file
Terminated
2022-08-30
Reason
The packaging sterile barrier does not comply with the applicable packaging standard for terminally sterilized medical devices. While the Instructions for Use (IFU) state Do not use if package has been previously opened or damaged , it is possible that the seal breach may go undetected by the user's naked eye.
Root cause
Under Investigation by firm
Action
Consignee Notification URGENT: MEDICAL DEVICE RECALL has been sent to US customers on 03-May-2018. For foreign customers, the Consignee Notification is in process. The Consignee letter informs customers of the situation and the required course of action which includes: 1. Cease use of the listed product 2. Account for number of product units used or discarded and return any units by completing the attached individualized Customer Confirmation form provided. Please follow the instructions described in the Customer Confirmation form. 3. Clinicians and their Patients who may have received the listed devices should monitor the patients condition to determine if there are any signs of infection. If this is the case, the patient should seek medical attention immediately to treat the infection. In addition, the Notification Letter should be provided to all those who need to be aware of this matter within your organization or to any organization where the potentially affected products may have been transferred. Fax the Customer Acknowledgement Response Form to Medtronic Neurosurgery at (805) 571-8787 or take a clear picture or scan of signed form and email it to rs.mnsfca@medtronic.com with "Magazine Recall" in the subject line.
Quantity
2616
Distribution
Domestic: AL, AZ, AR, CA, CO, DC, FL, GA, ID, IL, IN, IA, KS, LA, MD, MI, MN, MS, MO, NB, NV, NJ, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI, and WY OUS: Australia, Germany, Kenya, Russian Federation, South Africa
Lot, serial or model codes
Product Number/CFN: CM8902 UDI-Device Identifier (GTIN/UPN): 20643169647200 Lot Numbers: 54984; 57188; 59558; 66312; and 67988
510(k) numbers
["K050044"]
PMA numbers
not on file
Product code
HBO
Firm FEI
3015531529
Firm city
Goleta
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register