Skip to the content

FDA Medical Device Recalls (openFDA)

Accuray Incorporated: Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems

Recall number
Z-3211-2018
Recalling firm
Accuray Incorporated
Product
Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems
Status
Terminated
Initiated
2018-08-31
Posted
2018-09-20
Terminated
2020-12-16
Reason
the Ball Cube II phantom calibration film was not manufactured within necessary dimensions and may introduce up to 0.5 mm of positional inaccuracy to the CyberKnife System
Root cause
Nonconforming Material/Component
Action
The firm, Accuray, disseminated their "Urgent Field Safety Notices" dated 08/31/2018 by email and letter to its customers. The letter described the product, problem and actions to be taken. The letter requested the customers cease use and quarantine the product for recovery and replacement by an Accuray Representative and complete and return the Acknowledgement Form via email at: astsauver@accuray.com or mail to: Accuray Incorporated, one Erdman place, Madison, WI 53719. If you have any questions, contact Vice President, Regulatory Affairs at 608-824-2800 or email: dbiank@accuray.com or the Regulatory Affairs Specialist at 408-716-4660 or jysingh@accuray.com.
Quantity
259 units
Distribution
Worldwide Distribution: United States (nationwide), and countries of: United Kingdom, Germany, France, Finland, Netherlands, Portugal, Hong Kong, Poland, Australia, United Arab Emirates, Thailand, and Japan.
Lot, serial or model codes
All units with the Ashland Brand distributed between August 1, 2017 and July 31, 2018
510(k) numbers
not on file
PMA numbers
not on file
Product code
IYE
Firm FEI
3003873069
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register