FDA Medical Device Recalls (openFDA)
Microcare Medical: Good Vibrations Ultrasonic Cleaner Solution, JorVet J0382G Ultrasonic Instrument Cleaning Solution, SpecClean Ultrasonic Cleaner Solution, Models: GVUS128-1, GVUS128-1EA, GVUS550-1, GVUS24-1, GVUS24-1EA, JGVUS128-1, VGVUSNP128-1, VGVUSNP128-1EA
- Recall number
- Z-3209-2024
- Recalling firm
- Microcare Medical
- Product
- Good Vibrations Ultrasonic Cleaner Solution, JorVet J0382G Ultrasonic Instrument Cleaning Solution, SpecClean Ultrasonic Cleaner Solution, Models: GVUS128-1, GVUS128-1EA, GVUS550-1, GVUS24-1, GVUS24-1EA, JGVUS128-1, VGVUSNP128-1, VGVUSNP128-1EA
- Status
- Open, Classified
- Initiated
- 2024-08-29
- Posted
- 2024-09-23
- Terminated
- not on file
- Reason
- Contaminated equipment and packaging caused Klebsiella oxytoca, Enterobacter asburiae, Enterobacter bugandensis, pathogenic and opportunistic bacteria, to occur in ultrasonic detergent, which caused product bloating/leaking, and if used during medical/dental procedures could result in urinary tract, respiratory, bloodstream infections; workers handling contaminated product/instruments also at risk
- Root cause
- Material/Component Contamination
- Action
- On 8/29/24, recall notices were mailed to customers who were asked to do the following: 1) This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. This recall should be carried out to the user level. 2) If your company repacked these medical devices into convenience kits or into other packaging, please contact your local FDA recall coordinator to evaluate the need for a new recall (https://www.fda.gov/safety/industry-guidance-recalls/ora-recallcoordinators). 3) Return partially-used or unused product to the recalling firm. 4) Complete and return the response form via email to qdc@microcare.com If you have any questions, contact the firm at 303-529-5937 or the email above.
- Quantity
- 1649
- Distribution
- Worldwide - US Nationwide distribution in the states of MN, GA, PA, WA, CO, NY, MD, CA, MI, NE, UT, OK, WI, NJ, IL, ME, CT, KS, TX, NC, TN, SC, MO and the countries of Canada.
- Lot, serial or model codes
- Lot/Model: 2331982/GVUS128-1EA, GVUS128-1, GVUS550-1; 2332204/GVUS128-1EA, JGVUS128-1, GVUS128-1, VGVUSNP128-1EA, VGVUSNP128-1; 2332393/GVUS128-1EA, VGVUSNP128-1, GVUS24-1EA, GVUS24-1, GVUS128-1 2432755/JGVUS128-1 2432989/GVUS24-1EA, VGVUSNP128-1EA, GVUS24-1, VGVUSNP128-1, JGVUS128-1 Model/UDI-DI: GVUS128-1/+H816GVUS128H, GVUS24-1/+H816GVUSU24%, GVUS550-1/+H816GVUS550G, VGVUS128-1/+H816VGVUSNP128A, JGVUS128-1/723754044532
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JCB
- Firm FEI
- 1720499
- Firm city
- Commerce City
- Firm state
- CO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28