Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes USA (HQ), Inc.: Synthes T1 Matrix Locking Cap of the Synthes Matrix System, Part Number 04.632.000. Intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation.

Recall number
Z-3207-2011
Recalling firm
Synthes USA (HQ), Inc.
Product
Synthes T1 Matrix Locking Cap of the Synthes Matrix System, Part Number 04.632.000. Intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation.
Status
Terminated
Initiated
2011-03-11
Posted
2011-09-15
Terminated
2014-12-18
Reason
Labeling Correction -- The locking cap was not able to be loosened, resulting in the breakage of the screwdriver or the rod to be cut and the assembly to be removed (screw, cap, and rod). This poses a potential risk of significant prolongation of OR time.
Root cause
Labeling Change Control
Action
The firm initiated this action on March 11, 2011 by issuing an Urgent: Medical Device Labeling Correction letter to all consignees. The letter identified the affected product and stated the reason for the correction. It also instructed the consignee to review and attach the supplementary technique page to any MATRIX Spine System Technique Guides that they may have had in their facility. Customers are to complete the Verification Section of the letter and return it to the firm via mail, fax, or e-mail. If customers have questions, they can call 610-719-5063 or their Synthes Spine Sales Consultant.
Quantity
not on file
Distribution
Nationwide Distribution -- AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KS, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WV, and WI.
Lot, serial or model codes
All lots
510(k) numbers
["K092929","K100952"]
PMA numbers
not on file
Product code
NKB
Firm FEI
3005180112
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register