Skip to the content

FDA Medical Device Recalls (openFDA)

Howmedica Osteonics Corp.: BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-800 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Recall number
Z-3202-2026
Recalling firm
Howmedica Osteonics Corp.
Product
BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-800 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Status
Open, Classified
Initiated
2026-07-30
Posted
2026-09-16
Terminated
not on file
Reason
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Root cause
Process change control
Action
Firm began notifying customers on July 30, 2026, via FedEx 2-day air. Customers were instructed to quarantine and discontinue use of the affected product. Affected product should be returned to the firm.
Quantity
12,156 US
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Lot, serial or model codes
C-M113-1-301 UDI-DI 07613327590074 Lot MH621A C-M113-5-600 No UDI Lot 194593 C-M063-5-800 No UDI Lot RX06358A
510(k) numbers
not on file
PMA numbers
not on file
Product code
LGE
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register