FDA Medical Device Recalls (openFDA)

AVID Medical, Inc.: Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Model Number: KRIV26-02.

Recall number
Z-3194-2026
Recalling firm
AVID Medical, Inc.
Product
Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Model Number: KRIV26-02.
Status
Open, Classified
Initiated
2026-07-22
Posted
2026-09-23
Terminated
not on file
Reason
Convenience kits containing BD ChloraPrep Clear - 1 mL Applicator, which may be contaminated with Aspergillus penicillioides.
Root cause
Packaging process control
Action
Consignees were initially notified on about 07/22/2026 via "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter. They were informed that product may exhibit an open or incomplete seal on the packaging of the applicator. Consignees were instructed to quarantine impacted product, discontinue use of the BD ChloraPrep Clear - 1 mL Applicators within the kits, add a warning label to instruct users to discard the impacted component, and notify customers if the kits had been further distributed. Consignees were notified again on about 08/20/2026 with an updated "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter. Consignees were informed that certain lots of the BD ChloraPrep Applicators are contaminated with Aspergillus penicillioides. Instructions included to discard impacted kits containing affected applicators and notify customers if the kits had been further distributed.
Quantity
288 kits (6 cases)
Distribution
US Nationwide distribution.
Lot, serial or model codes
Kit Model Number: KRIV26-02. UDI-DI: 10809160460481. Lot Number: 1658437. Expiration Date: 1/31/2027
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRO
Firm FEI
1047429
Firm city
Toano
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register