Datex-Ohmeda, Inc.: Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

Recall number
Z-3191-2026
Recalling firm
Datex-Ohmeda, Inc.
Product
Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
Status
Open, Classified
Initiated
2026-07-30
Posted
2026-09-16
Terminated
not on file
Reason
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Root cause
Nonconforming Material/Component
Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/30/26 was sent to customers. Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use your Anesthesia system by following the instruction below: 1. Prior to each use, complete preoperative checkout and tests in accordance with the instructions in the User Reference Manual (URM). 2. When Alternate O2 control is enabled, the clinician should adjust the Alternate O2 flow as appropriate for the case. Use the Alternate O2 control knob to adjust the O2 flow. 3. When Alternate O2 control is enabled, flow from the electronic mixer is stopped and the agent concentration is set to off. Two situations are possible regarding the Agent quick key: a. The Agent quick key will be blank on the display if agent delivery is unavailable. Supplement with or transition to intravenous anesthetics as needed. b. The Agent quick key will be present on the display if agent delivery can be resumed. Press the Agent quick key to set the desired concentration and press the key again to confirm. 4. Please evaluate existing service stock for the Field Replaceable Unit (FRU) spare parts listed below and dispose of all affected FRU parts according to your facility procedures. Please ensure all potential staff in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 34147 Digital Response Form or print, fill out manually, scan, and email to recall.34147@gehealthcare.com. Product Correction GE HealthCare will inspect and correct as necessary all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. Contact Information If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity
7 units
Distribution
Worldwide - US Nationwide distribution.
Lot, serial or model codes
Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443
510(k) numbers
["K172045"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
2112667
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register