Skip to the content

FDA Medical Device Recalls (openFDA)

Swissray Medical: ddR Formula B X-ray System, Model ddR Formula B X-ray system used for imaging

Recall number
Z-3191-2018
Recalling firm
Swissray Medical
Product
ddR Formula B X-ray System, Model ddR Formula B X-ray system used for imaging
Status
Open, Classified
Initiated
2016-11-18
Posted
2018-09-18
Terminated
not on file
Reason
Potential for bucky (the part that holds the grid and is moveable to position the patient) moving unexpectedly.
Root cause
Device Design
Action
The firm, Swissray Medical, issued official communication to customers as a Action Notification Report/work Instructions on November 18 2016. The Work Instructions state the affected devices, the problem symptoms, causes of the problem, remedy, any tools required, and the step-by-step corrective actions. The manufacturer, Swissray Medical AG, distributed the Work Instructions for the FCO and the Action Notification Report Form to all distributors and service entities who would be performing corrective actions in the field. The Swissray Field Engineer then presented the Report to the customer (end user facility) and communicated with the customer in-person while performing the corrective action on-site. When corrective actions were completed, both the customer and service personnel then sign the Report to confirm that the customer was informed about the corrective action and that corrective action was completed. The Report Form states the title and unique designation of the corrective action, whether it was mandatory by the manufacturer or was an optional product improvement, and the unique document ID number of the Work Instructions for the FCO. The customers were instructed to Complete the FCO Action Notification Report and send to: Swissray Medical AG expert Center Turbistrasse 25-27 CH-6280 Hochdorf Switzerland Or scan and email to service@swissray.com with FCO in the subject line. If you have any questions, contact QA/RA Coordinator at 908-372-6408 or email: Deniz.Kortan@swissray.com.
Quantity
27 devices in US
Distribution
Worldwide Distribution: US (Nationwide) and to countries of: Austria, Dubai, China. Germany, Libya, Malaysia, Russia, and South Africa.
Lot, serial or model codes
Serial Numbers within U.S.A commerce - 1302002, 1302003, 1302005, 1302009, 1302011, 1402001, 1402002, 1402006, 1402007, 1402008, 1402009, 1402011, 1402012, 1402015, 1502001, 1502002, 1502004, 1502005, 1502006, 1502007, 1602001, 1602002, 1602003, 1602005, 1602006, 1602007 & 1602008   Serial Numbers Outside of U.S.A commerce - 1402005, 1402017, 1502009, 1502010, 1502011, 1302004, 1302006, 1302007, 1302008, 1302010, 1402003, 1402004, 1402010, 1402013, 1402014, 1402016, 1502008 & 1602004
510(k) numbers
["K052943"]
PMA numbers
not on file
Product code
KPR
Firm FEI
1000469975
Firm city
Hochdorf
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register