FDA Medical Device Recalls (openFDA)
Stryker Sustainability Solutions: REF:B5ST Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 5X75MM
- Recall number
- Z-3187-2026
- Recalling firm
- Stryker Sustainability Solutions
- Product
- REF:B5ST Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 5X75MM
- Status
- Open, Classified
- Initiated
- 2026-08-19
- Posted
- 2026-09-16
- Terminated
- not on file
- Reason
- Reusable access port system may crack or seperate during use.
- Root cause
- Device Design
- Action
- On August 19, 2026, Stryker Sustainability Solutions (SSS)issued a Urgent: Medical Device recall notice to affected consignees via sales representative and US Mail. SSS ask consignees to take the following actions: 1. Inform individuals within your organization who need to be aware of this action. 2. Immediately check your internal inventory to locate the affected product. Segregate the affected products, quarantine and discontinue use. 3. Complete the Business Reply Form or use the QR Code on Attachment A. Response is required. a.If you have indicated that product remains in your inventory, a prepaid shipping label will be issued for the return of the product. Credit will be issued to your facility for products returned upon receipt. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility.
- Quantity
- 9072 UNITS
- Distribution
- US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina , South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming and the countries of Israel and Canada.
- Lot, serial or model codes
- UDI:(01)00885825014254/ Lots: 16028740 16083100 16098933 16106547 16130667 16160384 16160418 16160460 16181332 16201793 16207031 16234052 16264116 16418941 16421915 16422882 16476685 16493231 16948345 16954673 16977948 16993151 17151380 17160201 17210355 17239624 17271015 17271027 17324927 17341521 17342144 17378894 17412852 17426084 17457260 17491569 17499741 17512459 17521310 17526338
- 510(k) numbers
- ["K201511"]
- PMA numbers
- not on file
- Product code
- NLM
- Firm FEI
- 3009350690
- Firm city
- Tempe
- Firm state
- AZ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28