Beckman Coulter, Inc.: Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro diagnostic instrument

Recall number
Z-3180-2024
Recalling firm
Beckman Coulter, Inc.
Product
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro diagnostic instrument
Status
Open, Classified
Initiated
2024-07-10
Posted
2024-09-19
Terminated
not on file
Reason
Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four complaints were filed for this event that led to Beckman Coulter finding undersized tubing in stock and initiated corrective actions. Use of the affected device may result in errors leading to delay of results.
Root cause
Nonconforming Material/Component
Action
Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 7/10/2024 via US mail and email. The notice explained the problem, impact, and requested the following actions be taken: If your laboratory observes either of these system eventscontact the Beckman Coulter Customer Technical Support Center at 1-800-854-3633 (US and Canada) or your local Beckman Coulter representative (outside US and Canada: a). Chassis.Pneumatics MachineControlStatus(Chassis maintain system air pressure failed. 1.The compressor has failed. 2.A large leak exists in the system) b). Pneumatics Air pressure under limit: ### PSI, Chassis.Pneumatics. WS04CompressedAir.32-35psi.PressureSensor. Beckman Coulter will schedule a time to replace the RSM pneumatic tubing.
Quantity
170 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czech Republic, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.
Lot, serial or model codes
UDI/DI 15099590230623, Serial Numbers: 300116, 300117, 300118, 300123, 300126, 300128, 300129, 300130, 300131, 300133, 300134, 300135, 300136, 300137, 300138, 300139, 300141, 300142, 300143, 300145, 300147, 300148, 300149, 300150, 300151, 300152, 300153, 300154, 300155, 300156, 300157, 300158, 300160, 300161, 300162, 300163, 300164, 300166, 300167, 300168, 300169, 300170, 300171, 300172, 300174, 300176, 300178, 300179, 300180, 300181, 300182, 300183, 300188, 300189, 300190, 300192, 300193, 300194, 300195, 300196, 300197, 300198, 300199, 300200, 300201, 300202, 300203, 300204, 300205, 300206, 300207, 300208, 300209, 300210, 300211, 300212, 300213, 300214, 300215, 300216, 300217, 300218, 300219, 300220, 300221, 300222, 300223, 300224, 300226, 300227, 300228, 300229, 300230, 300231, 300232, 300233, 300234, 300235, 300236, 300237, 300238, 300239, 300240, 300241, 300242, 300243, 300244, 300245, 300246, 300247, 300248, 300249, 300251, 300252, 300253, 300254, 300255, 300256, 300257, 300258, 300260, 300261, 300263, 300264, 300266, 300269, 300270, 300271, 300272, 300273, 300274, 300275, 300276, 300277, 300280, 300281, 300282, 300284, 300285, 300286, 300287, 300291, 300292, 300293, 300294, 300295, 300297, 300307, 300308, 300310, 300312, 300313, 300314, 300315, 300316, 300317, 300319, 300320, 300321, 300322, 300323, 300324, 300325, 300326, 300327, 300328, 300329, 300330, 300332, 300333
510(k) numbers
["K221225"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2122870
Firm city
Chaska
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register