restor3d Inc.: Kinos Axiom Total Ankle System with Axiom PSR, Flat-cut (Cut Guide), REF: 5831-9006

Recall number
Z-3178-2026
Recalling firm
restor3d Inc.
Product
Kinos Axiom Total Ankle System with Axiom PSR, Flat-cut (Cut Guide), REF: 5831-9006
Status
Open, Classified
Initiated
2026-07-27
Posted
2026-09-16
Terminated
not on file
Reason
Ankle implantation systems were packaged with incorrect patient-specific cutting guides, which may result in: delayed surgery, the need to use standard instrumentation, reduced fixation or poor alignment of implants, implant loosening/migration, or need for revision surgery.
Root cause
Labeling Change Control
Action
On 7/28/2026, recall notices were emailed to customers who were asked to do the following: 1) Segregate and return affected devices. 2) Complete and return the response form. If additional questions, contact firm at 984-201-2310
Quantity
2
Distribution
US Nationwide distribution in the states of FL, CO.
Lot, serial or model codes
System Serial Numbers: 18169, 18059. REF: 5831-9006, Lot: D-788, UDI-DI: 00840097509945.
510(k) numbers
["K223326"]
PMA numbers
not on file
Product code
OYK
Firm FEI
3014833750
Firm city
Durham
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register