Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Access Systems: BardPort Titanium Implanted Port. Bard Access Systems. Assembled in Mexico. A totally implantable vascular access device designed to provide long term repeated access to the vascular system

Recall number
Z-3173-2011
Recalling firm
Bard Access Systems
Product
BardPort Titanium Implanted Port. Bard Access Systems. Assembled in Mexico. A totally implantable vascular access device designed to provide long term repeated access to the vascular system
Status
Terminated
Initiated
2011-08-30
Posted
2011-09-08
Terminated
2012-03-30
Reason
Implanted port has discrepant product information on package labeling.
Root cause
Labeling Change Control
Action
Bard Access System, Inc. sent a "Label Discrepancy Notification dated August 30, 2011 to all affected customers. The letter included affected product and informed them of the labeling discrepancy. Customers could either use the device or return it to Bard for replacement. For information on this recall call (801) 595-5568.
Quantity
1375 units
Distribution
Worldwide Distribution: Nationwide Distribution including the states of; AL, AR, AZ, CA, CT, DC, FL, GA, HI, IL, KY, LA, MA, MD, MI, MO, NJ, NV, NY, OK, PA, RI, SC, TN, TX, UT, VA, WI, and WV; and the countries of Canada and Belgium.
Lot, serial or model codes
Prod Number: 0602290. Lot Numbers: RESF1061, RESI0209, RESI0672, RESI1072, RESK0055, RESK0806, RETA0584, RETC0929, RETF0739, RETH0830.
510(k) numbers
["K050310"]
PMA numbers
not on file
Product code
not on file
Firm FEI
3006260740
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register