Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare: Centricity PACS Server V7, intended for managing digital medical images and associated data, All versions

Recall number
Z-3169-2026
Recalling firm
GE Healthcare
Product
Centricity PACS Server V7, intended for managing digital medical images and associated data, All versions
Status
Open, Classified
Initiated
2026-07-31
Posted
2026-09-15
Terminated
not on file
Reason
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
Root cause
Under Investigation by firm
Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/31/26 was sent to customers. Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use your systems by following the instructions below. 1. Do not perform any new inbound data (DICOM) migration projects. Please contact GE HealthCare before starting any new migration projects. 2. If using Enterprise Archive with Core Services Hub/Workflow Core Services, when reviewing or exporting prior imaging studies after migration to Enterprise Archive, verify that the study and images within the study are associated with the correct patient as follows: At the Study level: " Compare name, date of birth, and patient identifiers only if burned-in or embedded in the image. Do not rely on patient information in the image overlays that can be toggled on and off. " If burned-in or embedded patient identifiers are not available, compare with prior studies based on patient anatomy, implants, or unique anatomical features. At the Image level: " Confirm that the images within the study do not belong to different patients based on acquisition date and time or other patient identifiers. Please maintain any previous source systems or migration servers pending correction of this issue by GE HealthCare. Please ensure that all potential users in your facility are made aware of this safety notification and the recommended actions above. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 85492 Digital Response Form or print, fill out manually, scan, and email to recall.85492@gehealthcare.com. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity
116 systems
Distribution
Worldwide - US Nationwide distribution.
Lot, serial or model codes
UDI-DI: 00195278882585; System IDs: 847NWCPVSW 847NWCPVSW2 NORTHWEST25206018 207795PVSWC3 207795PVSWC1 207795PVSWC2 604520GEPACS 613548IMSA 613548IMSB 613548TESTIMS 613548GEPACS EG1055PC01 B4191728 C4191734 D4191718 E4191720 tobecreated HHEI02CS01 UC1989CS01 CPACS Testsystem Not Available HCIT5592143PACS JL1758 JD1266 JL1760 JAS004 JL1762 JAS042 JL1757 JL1761 JL1759 KW1018PP01 ALRAZIIMS LB1279PC02 BMALTA25 PIV-PIMS1 SA2591PP01 SA2039PP01 SA2315PP12 KAMCJIMS2 SA1208PP01 SA2000PP01 SA1006PP07 KFHIMS2 KAHIMS2 MKNSHIMS2 ZA2895PACS02 FEG85600 FEG85608 FEG34405 FEG34406 FEB92518 FEB92519 FEE07A22 FEE07A23 FEG85700 FEG85704 FE410911 FE410912 IMSTEST TBC DE619840 DE619846 DEM88511 DEM88512 DEM87116 DEM87121 DEI12311 DEI12312 DEP00203 20034126 NE465120 NE670325 NE673137 NE300609 NE219643 FE816139 FE816140 FE8161102 DPP03300 DPP03301 DPP03302 DPP03303 DPP03305 FEG51800 FEF70522 FEF70523 MEX61704 MEX67314 GPX24300 NPV73804 25277590 NE203141 NE740521 NE216225 NE487826 MEX67605 NE723605 NEV07027 NE748614 NEV13618 NEV06838 NE304907 NE223245 NE419140 NEV44001 NPV84521 NEV44017 NEV44018 NEV44034 NE741437 NEV35826 NE688826 GEE34022 06002PAC37 06051PAC01
510(k) numbers
["K043415","K110875"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3004526608
Firm city
Chicago
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register