Skip to the content

FDA Medical Device Recalls (openFDA)

Btt Health Gmbh: Marodyne-LiV, part number 3045-133, Low intensity vibration exerciser supporting osteoporosis treatment

Recall number
Z-3162-2024
Recalling firm
Btt Health Gmbh
Product
Marodyne-LiV, part number 3045-133, Low intensity vibration exerciser supporting osteoporosis treatment
Status
Open, Classified
Initiated
2024-08-01
Posted
2024-09-17
Terminated
not on file
Reason
Some power supply units in Austria and Germany had a voltage drop in the millisecond range, which led to a cancellation of the treatment and subsequent restart of the medical device. The treatment is only restarted by a step-off/step-on (step-on for activation).
Root cause
Component design/selection
Action
Btt notified its distributors on 08/01/2024 via email. The notice explained the problem with the device, impact the the user, and were directed to notify their customers. The power supplies will be replaced.
Quantity
62 units
Distribution
US Nationwide distribution in the states of Florida.
Lot, serial or model codes
serial numbers: 010426065891001421S05327911240506, 010426065891001421S05328011240506, 010426065891001421S05328111240506, 010426065891001421S05328211240506, 010426065891001421S05328311240506, 010426065891001421S05328411240506, 010426065891001421S05328511240506, 010426065891001421S05328611240506, 010426065891001421S05328711240506, 010426065891001421S05328811240506, 010426065891001421S05328911240506, 010426065891001421S05329011240506, 010426065891001421S05321911240430, 010426065891001421S05322011240430, 010426065891001421S05322111240430, 010426065891001421S05322211240430, 010426065891001421S05322311240430, 010426065891001421S05322411240430, 010426065891001421S05322511240430, 010426065891001421S05322611240430, 010426065891000721S05336411240521, 010426065891000721S05336511240521, 010426065891000721S05336611240521, 010426065891000721S05336711240521, 010426065891000721S05336811240521, 010426065891000721S05336911240521, 010426065891000721S05337011240521, 010426065891000721S05337111240521, 010426065891000721S05337211240521, 010426065891000721S05337311240521, 010426065891000721S05337411240521, 010426065891000721S05337511240521, 010426065891000721S05337611240521, 010426065891000721S05337711240521, 010426065891000721S05337811240521, 010426065891000721S05337911240521, 010426065891000721S05338011240521, 010426065891000721S05338111240521, 010426065891000721S05338211240521, 010426065891000721S05338311240521, 010426065891000721S05338411240521, 010426065891000721S05338511240521, 010426065891000721S05338611240521, 010426065891000721S05338711240521, 010426065891000721S05338811240521, 010426065891000721S05338911240521, 010426065891000721S05339011240521, 010426065891000721S05339111240521, 010426065891000721S05339211240521, 010426065891000721S05339311240521, 010426065891000721S05339411240521, 010426065891000721S05339511240521, 010426065891000721S05339611240521, 010426065891000721S05339711240521, 010426065891000721S05339811240521, 010426065891000721S05339911240521, 010426065891000721S05340011240521, 010426065891000721S05340111240521, 010426065891000721S05340211240521, 010426065891000721S05340311240521, 010426065891000721S05340411240521, 010426065891000721S05340511240521
510(k) numbers
not on file
PMA numbers
not on file
Product code
BXB
Firm FEI
3013515165
Firm city
Inning
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register