Skip to the content

FDA Medical Device Recalls (openFDA)

Orthofix, Inc.: Ratcheting Torque Limiting Handle, Catalog Number 52-1012, Orthofix Spinal Implants, Firebird Spinal Fixation System, Firebird Spinal Fixation System DDD Instrument Case. Product is intended for use with the Set Screw Driver (52-1061) and Adjustable Counter Torque Wrench (52-1265) to perform the final tightening of set screws in a spinal implant construct.

Recall number
Z-3162-2011
Recalling firm
Orthofix, Inc.
Product
Ratcheting Torque Limiting Handle, Catalog Number 52-1012, Orthofix Spinal Implants, Firebird Spinal Fixation System, Firebird Spinal Fixation System DDD Instrument Case. Product is intended for use with the Set Screw Driver (52-1061) and Adjustable Counter Torque Wrench (52-1265) to perform the final tightening of set screws in a spinal implant construct.
Status
Terminated
Initiated
2011-04-01
Posted
2011-09-07
Terminated
2012-03-23
Reason
Upon reaching torque limit, there is a possibility the product may jam or exhibit a ratchet direction switch from forward to reverse.
Root cause
Device Design
Action
Orthofix, Inc. sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated April 1, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to identify and remove the affected product from use. Customers were advised to use the combination of the Ratching R Handle and the Set Screw Driver if they experience problems with the Ratching Torque Limiting Handle. A Tracking and Verification form was attached for customers to complete and return to the Orthofix Inc. Regulatory Affairs Department via fax at 214-937-3322. Contact your Account Services Representative at 888-298-5700 for questions regarding the removal and return of the affected product.
Quantity
287 units
Distribution
Worldwide Distribution-USA (nationwide) and Puerto Rico and the countries of Australia and Germany.
Lot, serial or model codes
All serial numbers.
510(k) numbers
["K092624"]
PMA numbers
not on file
Product code
NKB
Firm FEI
2183449
Firm city
Lewisville
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register