FDA Medical Device Recalls (openFDA)
ICU Medical, Inc.: Oncology sets with drip chambers
- Recall number
- Z-3159-2026
- Recalling firm
- ICU Medical, Inc.
- Product
- Oncology sets with drip chambers
- Status
- Open, Classified
- Initiated
- 2026-05-12
- Posted
- 2026-09-11
- Terminated
- not on file
- Reason
- Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.
- Root cause
- Process control
- Action
- On 5/12/2026, the firm posted a Information Regarding Reported Black Specks in Certain IV Tubing Set Drip Chambers from ICU Medical document on their website. On 5/18/2026, updated the notices were sent. On 7/28/2026, correction notices were sent to customers asking them to do the following: 1) Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used or further distributed to another location, please ensure that this communication is delivered there. 2) Complete and return the attached Customer Response Form to MarketAction@mailac.custhelp.com 3) Continue using products as intended and per facility protocol. 4) Report any product quality concerns to ICU Medical Complaint Management. Firm contacts: GlobalComplaints@icumed.com 1-866-261-8806. customerservice@icumed.com 1-800-258-5361
- Quantity
- 28,550
- Distribution
- Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and the countries of PK, PH, MY, HK, IN, DE, CR.
- Lot, serial or model codes
- All products manufactured on or before 7/4/2026. UDI-DI: 10887709076768, 10887709088105
- 510(k) numbers
- ["K131549","K173477"]
- PMA numbers
- not on file
- Product code
- ONB
- Firm FEI
- 2025816
- Firm city
- San Clemente
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28