Skip to the content

FDA Medical Device Recalls (openFDA)

FUJIFILM Healthcare Americas Corporation: The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.

Recall number
Z-3158-2024
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.
Status
Open, Classified
Initiated
2023-11-23
Posted
2024-09-25
Terminated
not on file
Reason
When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.
Root cause
Radiation Control for Health and Safety Act
Action
Customer and Dealer notification will occur by means of letters alerting customers and dealers to the potential for overexposure to children if the conditions are not checked after the key switch is turned on. The notices also include the plan for the affected mobile x-ray systems. The system will be upgraded, free of charge, with updated software so that the exposure values set by the user are reflected regardless of the state of the key switch.
Quantity
130
Distribution
US Nationwide and Worldwide
Lot, serial or model codes
All serial numbers shipped from FUJIFILM
510(k) numbers
["K170858"]
PMA numbers
not on file
Product code
IZL
Firm FEI
1000513161
Firm city
Lexington
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register