FDA Medical Device Recalls (openFDA)
FUJIFILM Healthcare Americas Corporation: The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.
- Recall number
- Z-3158-2024
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.
- Status
- Open, Classified
- Initiated
- 2023-11-23
- Posted
- 2024-09-25
- Terminated
- not on file
- Reason
- When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.
- Root cause
- Radiation Control for Health and Safety Act
- Action
- Customer and Dealer notification will occur by means of letters alerting customers and dealers to the potential for overexposure to children if the conditions are not checked after the key switch is turned on. The notices also include the plan for the affected mobile x-ray systems. The system will be upgraded, free of charge, with updated software so that the exposure values set by the user are reflected regardless of the state of the key switch.
- Quantity
- 130
- Distribution
- US Nationwide and Worldwide
- Lot, serial or model codes
- All serial numbers shipped from FUJIFILM
- 510(k) numbers
- ["K170858"]
- PMA numbers
- not on file
- Product code
- IZL
- Firm FEI
- 1000513161
- Firm city
- Lexington
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28