Skip to the content

FDA Medical Device Recalls (openFDA)

Mazor Robotics Ltd: Mazor X. Model Number REF TPL0059; stereotaxic orthopedic instrument

Recall number
Z-3156-2024
Recalling firm
Mazor Robotics Ltd
Product
Mazor X. Model Number REF TPL0059; stereotaxic orthopedic instrument
Status
Open, Classified
Initiated
2024-07-17
Posted
2024-09-16
Terminated
not on file
Reason
Software anomalies which can result in two separate issues: A lagging screen display during a navigated procedure; or the potential for a mismatch between the navigated screw and the screw information presented on the screen.
Root cause
Device Design
Action
Medtronic/Mazor Robotics, Ltd. issued an URGENT: Medical Device Software Correction to its consignees on 07/17/2024 via 2-day tracked mailing. The notice explained the issue, potential risk, and provided a workaround until the software update is installed. Medtronic will schedule a time to perform a software update that will resolve these anomalies. Medtronic field service engineers will update all Mazor X systems from version 5.1.1 to version 5.1.2 or later. Questions: contact Medtronic Technical Services at 1-888-826-5603 or rs.navtechsupport@medtronic.com .
Quantity
94 devices
Distribution
US Nationwide distribution in PR.
Lot, serial or model codes
UDI/DI 07290109180465, Lot Numbers: 400-38-122016, 400-44-022017, 400-67-062017, 400-69-062017; UDI/DI 07290109181158, Lot Numbers: 4001200218, 4001970419, 4001011117, 4001041117, 4001061117, 4001320518, 4001330618, 4001390818, 4001430918, 4001631218, 4001690119, 4001710119, 4001780219, 4001790219, 4001800219, 4001960419, 4001980419, 4002050419R, 4002170619, 4002230819, 4002240819, 4002310919, 4002511119, 4002531119, 4002541119, 4002591119, 4002641219, 4002661219, 4002850220, 4002860220R, 4002900220, 4003020320, 4003030320, 4003180320, 4003190320, 4003430920, 4006951223, 4006901223, 4006890923, 4006860923, 4006830923, 4006680723, 4006170323, 4006090223, 4006050223, 4006040223, 4005210922, 4005770123, 4005720123, 4005431122, 4005750123, 4006130323, 4004091121; UDI/DI 07290109184098, Lot Numbers: 4003840821, 4003860821, 4004151121, 4004231221, 4004290122, 4004310122; UDI/DI 07290115751376, Lot Numbers: 4004860522, 4005180822, 4004670322, 4004700322, 4004870522, 4005080722, 4005130822, 4005340922, 4005401122, 4005421122, 4005451122, 4005481122, 4005491122, 4005591222, 4005760123, 4005790123, 4005870123, 4005900223, 4005910223, 4005940223, 4005980223, 4006030223, 4006140323, 4006230423, 4006330523, 4006630723, 4006660723, 4006670723, 4006740823, 4006790823, 4006800823
510(k) numbers
["K230064"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3005075696
Firm city
Caesarea
Firm state
not on file
Firm country
Israel

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register