Skip to the content

FDA Medical Device Recalls (openFDA)

Tornier, Inc: stryker TORNIER PERFORM HUMERAL SYSTEM - STEMLESS US Anatomic Nucleus, REF DWM003A; anatomic total arthroplasty of the shoulder

Recall number
Z-3155-2024
Recalling firm
Tornier, Inc
Product
stryker TORNIER PERFORM HUMERAL SYSTEM - STEMLESS US Anatomic Nucleus, REF DWM003A; anatomic total arthroplasty of the shoulder
Status
Open, Classified
Initiated
2024-08-29
Posted
2024-09-16
Terminated
not on file
Reason
One lot of Perform Humeral Nucleus may exceed the specification for bacterial endotoxin load (max 20 EU/device).
Root cause
Under Investigation by firm
Action
Stryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 08/29/2024 via mail and email. The notice explained the issue, potential risk, and requested the following actions be taken: identify and isolate all affected product to prevent accidental use, and complete and return the response form. Upon receipt, Stryker will contact the consignee and arrange for the return of the product. Distributor were directed to notify their customers. Stryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 08/29/2024 via mail and email. The notice explained the issue, potential risk, and requested the following actions be taken: identify and isolate all affected product to prevent accidental use, and complete and return the response form. Upon receipt, Stryker will contact the consignee and arrange for the return of the product. Distributors were directed to notify their customers. For questions or concerns, please contact memphis.fieldaction@stryker.com.
Quantity
5 units
Distribution
US Nationwide distribution in the states of IL, WI, AZ, NV, SD.
Lot, serial or model codes
UDI/DI 00846832092192, Lot number 3268BA
510(k) numbers
["K220418"]
PMA numbers
not on file
Product code
PKC
Firm FEI
3000718467
Firm city
Bloomington
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register