Medtronic Navigation, Inc.: Spine Referencing Instrumentation, Set, Contains Spinous Process Clamps, Short and Tall When used with a Medtronic StealthStation Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery. The devices are intended to be reusable. The spine referencing clamps provide bony fixation through two styles of clamps; a single spinous process clamp and a double spinous process clamp.

Recall number
Z-3143-2017
Recalling firm
Medtronic Navigation, Inc.
Product
Spine Referencing Instrumentation, Set, Contains Spinous Process Clamps, Short and Tall When used with a Medtronic StealthStation Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery. The devices are intended to be reusable. The spine referencing clamps provide bony fixation through two styles of clamps; a single spinous process clamp and a double spinous process clamp.
Status
Terminated
Initiated
2017-06-29
Posted
not on file
Terminated
2023-07-24
Reason
Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.
Root cause
Device Design
Action
Medtronic sent an Urgent Medical Device Correction letter dated June 27, 2017, to all affected consignees. The letter notifies consignees of the product correction and requests that practitioners follow the attached use instructions for devices. Practitioners are to identify product damage before each procedure by opening and closing the device jaw with the T-handle. It is instructed that practitioners not continue turning the screw if resistance is met, as this may break the washer. Clamps are not to be used if the jaws remain closed when attempting to open a clamp or if the clamp screw back is out of the clamp. It is asked that this check is done to all inventory and any damaged product is to be reported to Medtronic at 1-800-595-9709 and quarantined. Revised product is available. Consignees are asked to complete and return the provided Consignee Response Form by email to the firm at RS.NavFCA@medtronic.com or by fax to 651-367-7075. Any questions can be directed to 1-800-595-9709.
Quantity
2,051 units total
Distribution
Worldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, AD, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, & WY. Foreign: Albania, Australia, Bangladesh, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Indonesia, Israel, Italy, Japan, Kenya, Latvia, Luxembourg, Malaysia, Martinique, Netherlands, New Zealand, Philippines, Poland, Portugal, Romania, Russia, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, & Vietnam.
Lot, serial or model codes
P/N 9734715, 9734716
510(k) numbers
["K131425"]
PMA numbers
not on file
Product code
OLO
Firm FEI
1000517638
Firm city
Louisville
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register