Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N

Recall number
Z-3137-2026
Recalling firm
Medline Industries, LP
Product
Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N
Status
Open, Classified
Initiated
2026-07-23
Posted
2026-09-11
Terminated
not on file
Reason
The kits contain a Retrograde Coronary Sinus Perfusion (RCSP) cannulae that has the potential for a sterile barrier breach. Sterility of the product cannot be assured. If a compromised pouch seal is not identified prior to use and the cannula is used, there is a potential risk of infection or other complications associated with the use of a non-sterile device, including hemolysis, foreign body reaction, thromboembolism, and/or organ dysfunction.
Root cause
Packaging process control
Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 7/23/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-121-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
135 units
Distribution
US Nationwide distribution in the state of CA.
Lot, serial or model codes
UDI-DI: 10198459445941 (each), 40198459445942 (case); Lot Numbers: 25HLA462, 25HLA916, 25JLA847, 25LLA298, 26CLA372
510(k) numbers
["K213481"]
PMA numbers
not on file
Product code
LRO
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register